81
Balance internal error
If this error appears persistently, contact the local A&D dealer.
Battery error for clock
The back-up battery power for the clock was lost completely.
After pressing one of the keys, adjust the date and time. Even if
the back-up battery power for the clock was lost and the balance
is on or in the standby state, the balance performance will be
normal. When this error appears frequently, repair is required.
Low battery error
Voltage of the power supply drops.
Confirm whether the correct AC adapter is used.
EC, E00 Communications error
A protocol error occurred in communications.
Confirm the format, baud rate and parity.
EC, E01 Undefined command error
An undefined command was received.
Confirm the command.
EC, E02 Not ready
A received command can not be processed.
e.g. The balance received a Q command, but not in the
weighing mode.
e.g. The balance received a Q command while processing a
RE-ZERO command.
Adjust the delay time to transmit a command.
EC, E03 Timeout error
If the timeout parameter is set to “
t-Up1
”, the balance did not
receive the next character of a command within the time limit of
one second.
Confirm the communication.
EC, E04 Excess characters error
The balance received excessive characters in a command.
Confirm the command.
EC, E06 Format error
A command includes incorrect data.
e.g. The data is numerically incorrect.
Confirm the command.
EC, E07 Parameter setting error
The received data exceeds the range that the balance can
accept.
Confirm the parameter range of the command.
20.3. Asking For Repair
If the balance needs service or repair, contact the local A&D dealer.
The balance is a precision instrument. Use much care when handling the balance and observe the
following when transporting the balance.
Use the original packing material.
Remove the weighing pan from the main unit