21
Using The Accumulation Function
Use the following keys to operate the accumulation function.
MODE
key .......... Displays the weighing data and the total value alternately each time it is
pressed. While the accumulation function is in use, the unit can not be
changed.
RE-ZERO
key ..... Sets the display to zero while the weighing data is displayed.
Deletes the total value while the total value is displayed.
key .......... Outputs and adds the weighing data while the weighing data is displayed.
Outputs the total value while the total value is displayed.
Step 1 Press the
RE-ZERO
key to zero the
display.
Step 2
Place a sample on the pan. The weight
value is displayed.
Step 3 Press the
key. The weight
value is added to the total and is
output.
The
accumulation
number at the upper
left of the display increases by one.
Step 4 Repeat steps 1 to 3, when
accumulating more data.
Step 5 Press the
MODE
key to display
the total value.
Outputting the value
Step 6 Press the
key to output the
total value.
Step 7 Press the
RE-ZERO
key to delete
the total value.
Notes And Displaying Or Outputting An Overloaded Total
The output format depends on the function table setting.
While the accumulation function is in use, the data memory
function is not available.
To disable the accumulation function, set the
"Accumulation function (
add
)" parameter to "
0
".
When the "Data number output (
d-no
)" parameter is set to
"
1
", the accumulation number will be output before the
weighing data.
Output ST,+012345.6
g
Accumulation
number
Accumulation
number
Output TOTAL(N=012)
ST,+012345.6
g
Accumulation value
Data number is above 99
Accumulation data is
overloaded.
Output TOTAL(N=--)
OL,+919
g
Summary of Contents for GF-30K
Page 7: ...5...
Page 95: ...93 20 1 External Dimensions Unit mm...
Page 102: ...100 MEMO...
Page 103: ......