23
7-2. Self Check Function With Response Adjustment For The GX-K Series
This function automatically updates the response adjustment by analyzing the influence of
the environment on the weighing data and also self-checks the balance performance using
the internal mass.
Step 1 Press and hold the
MODE
key until
RESPONSE
is
displayed, and then release the key.
Step 2 The balance automatically starts to check the balance
performance and sets the response characteristic.
Caution Do not allow vibration or drafts to affect the
balance during adjustment.
Step 3 After automatic adjustment, the balance displays the updated
response indicator and returns to the weighing mode.
The response indicator remains displayed for a while.
Example of display
MID.
and
OK
: The example above indicates that the result of the self check
is good and MID. is selected as the response rate..
Note
If improper performance is found in the self check, the balance displays
CH no
.
Contact the local A&D dealer for repair.
If the automatic response adjustment fails, the balance displays
CH ng
. Check the
ambient conditions such as breeze and vibration, also check the weighing pan. Then,
perform the adjustment again. To return to the weighing mode, press the
CAL
key.
Advise
If the automatic response adjustment is not helpful, try to refine it using the "7-3. Self Check
Function For GF-K Series".
7-3. Self Check Function For GF-K Series
This function manually updates the response adjustment by analyzing the influence of the
environment on the weighing data.
Step 1 Press and hold the
MODE
key until
RESPONSE
is displayed.
And then, press the
MODE
key again quickly.
Step 2 Press
the
MODE
key to select a weighing speed.
Either
FAST
,
MID.
or
SLOW
can be selected.
Step 3 After a few seconds of inactivity the balance displays
end
.
Then, it returns to the weighing mode and displays the
updated response indicator.
The response indicator remains displayed for a while.
Note
If improper performance is found in the self check, the balance displays
CH no
.
Contact the local A&D dealer for repair.