37
10-3. Description Of The Class "Environment, Display"
Condition (
Cond
)
Cond 0
This parameter is for sensitive response to the fluctuation of a weight value. Used
for powder target weighing, weighing a very light sample or when quick response
weighing is required. After setting, the balance displays
FAST
.
Cond 2
This parameter is for stable weighing with slow response. Used to prevent a
weight value from drifting due to vibration or drafts. After setting, the balance
displays
SLOW
.
Notes In automatic response adjustment, the weighing speed is selected
automatically.
With "Hold function (
Hold
)" set to "ON (
1
)", this item is used to set the
averaging time.
Stability band width (
5t-b
)
This item controls the width to regard a weight value as a stable value. When the fluctuation
per second is less than the parameter, the balance displays the stabilization indicator and
outputs or stores the data. The parameter influences the "Auto print mode"
5t-b
0
This parameter is used for sensitive response of the stabilization indicator. Used
for exact weighing.
5t-b
2
This parameter ignores slight fluctuation of a weight value. Used to prevent a
weight value from drifting due to vibration or drafts.
Note
With "Hold function (
Hold
)" set to "ON (
1
)", this item is used to set the
stabilization range.
Hold function (
Hold
) (Animal weighing mode)
This function is used to weigh a moving object such as an animal. When the weighing data
is over the weighing range from zero and the display fluctuation is within the stabilization
range for a fixed period of averaging time, the processing indicator illuminates and the
balance displays the average weight of the animal. When the animal is removed from the
weighing pan, the display returns to zero automatically. This function is available only when
the hold function parameter is set to "
1
" (the animal mode indicator
ANIMAL
illuminates)
and any weighing unit other than the counting mode is selected. The stabilization range and
averaging time are set in "Condition (
Cond
)" and "Stability band width (
5t-b
)".
Weighing range
Averaging time
Stable range
GX-8K GX-8K2 GX-10K
Cond 0
5t-b 0
GF-8K GF-8K2 GF-10K
2g
Cond 1
5t-b 1
Cond 2
2 sec. Faster
4 sec.
8 sec.More accurate
5t-b 2
Small
Big
GX-12K GX-20K
GX-30K GX-32K
GF-12K GF-20K
GF-30K GF-32K
20g
Summary of Contents for GF-30K
Page 7: ...5...
Page 95: ...93 20 1 External Dimensions Unit mm...
Page 102: ...100 MEMO...
Page 103: ......
Page 104: ......