40
Interval memory mode
The weighing data is periodically stored in memory.
Example
For periodical weighing without a personal computer command and
to output all of the data, to a computer, at one time.
The GX-K series can use time and date with "Time/Date output (
5-td
)".
Required
setting
dout
prt 3
Interval memory mode
dout
data 2
Data memory function is used
dout
int
Interval
time
Optional
setting
dout
5-td1
,
2
, or
3
Adds the time and date.
10-5. Description Of The Item "Data Format"
A&D standard format
5if type 0
This format is used when the peripheral equipment can receive the A&D format.
If an AD-8121B is used, set the printer to MODE 1 or 2.
This format consists of fifteen characters excluding the terminator.
A header of two characters indicates the balance condition.
The polarity sign is placed before the data with the leading zeros. If the data is zero, the plus
sign is applied.
The unit, consisting of three characters, follows the data.
DP (Dump print) format
5if type 1
This format is used when the peripheral equipment can not receive the A&D format.
If an AD-8121B is used, set the printer to MODE 3.
This format consists of sixteen characters excluding the terminator.
A header of two characters indicates the balance condition. No overload header is used.
The polarity sign is placed before the data, with spaces in place of leading zeros, if the data
is not zero or overloaded.
The unit, consisting of three characters, follows the data.
Stable header
Stable header of counting mode
Unstable header
Overload header
Header Data
Unit
Terminator
Stable header
Stable header of counting mode
Unstable header
Header Data
Unit
Terminator