53
11-2. GLP Report
Set the following parameters to output the report.
To print the report, set the "GLP output (
info
)" parameter to "
1
" and use MODE 3 of the
AD-8121B. Refer to "16-2. Connection to the AD-8121B Printer" for details on using the
printer.
To output the report to a personal computer using the RS-232C interface, set the "GLP
output (
info
)" parameter to "
2
".
If the time and date are not correct, set the correct time and date in "Clock (
Cl adj
)" of the
function table.
Notes
For operational details about calibration and calibration test, refer to "8. Calibration".
The GF-K series does not output time and date to the general format.
Calibration report using the internal mass
Setting of "
info 1
"
Setting of "
info 2
"
AD-8121 printer format
General format
Space, ASCII 20h
<TERM>
Terminator, CR , LF or CR
CR
Carriage return, ASCII 0Dh
LF
Line feed, ASCII 0Ah
Manufacturer
Model
Serial number
ID number
Date
Time
Calibration type
Signature
===========
A
=
&
=
D
MODEL
=====
GX-30K
S/N
=====
01234567
ID
=======
ABCDEFG
DATE
======
2004/12/31
TIME
========
12:34:56
CALIBRATED(INT.)
SIGNATURE
- - - - - - - -
~~~~~~~~~~~
A
~
&
~
D
<TERM>
MODEL
~~~~~
GX-30K
<TERM>
S/N
~~~~~
01234567
<TERM>
ID
~~~~~~~
ABCDEFG
<TERM>
DATE
<TERM>
~~~~~~
2004/12/31
<TERM>
TIME
<TERM>
~~~~~~~~
12:34:56
<TERM>
CALIBRATED(INT.)
<TERM>
SIGNATURE
<TERM>
<TERM>
<TERM>
----------------
<TERM>
<TERM>
<TERM>
Summary of Contents for GF-30K
Page 7: ...5...
Page 95: ...93 20 1 External Dimensions Unit mm...
Page 102: ...100 MEMO...
Page 103: ......
Page 104: ......