58
"
fUl
" indicates that memory is full or the memory capacity has been reached. More data can
not be stored unless the memory data is deleted.
Automatic self calibration can not be used while the interval memory mode is active.
The following commands can not be used during data storage.
Q
Query command for weighing data.
S
Query command for stable weighing data.
SI
Query command for weighing data.
SIR
Query command for continuous weighing data.
Setting the function table
Parameter settings for each output mode are as follows:
Item
Mode
Data output
mode
Auto print polarity,
difference
Data memory
function
Interval time
Key mode
prt 0
Not used
data 2
Auto print mode A
prt 1
data 2
Auto print mode B
prt 2
ap-a 0-2
ap-b 0-2
data 2
Not used
Interval memory mode
prt 3
Not used
data 2
int 0-8
Set each item for GX-K series, depending on the situation, as follows:
No
d-no 0
No
5-td 0
Data number
Yes
d-no 1
Time only
5-td 1
No
5-id 0
Date only
5-td 2
ID number
Yes
5-id 1
Time and date
Both
5-td 3
When
5-td 1
,
2
or
3
is selected, the amount of data to be stored is 100 sets.
Set each item for GF-K series, depending on the situation, as follows:
No
d-no 0
Data number
Yes
d-no 1
No
5-id 0
ID number
Yes
5-id 1
GF-K series does not use time and date.
Enabling the data memory function
Step 1
Press and hold the
SAMPLE
key until
ba5fnc
is displayed, then release the key.
Step 2 Press
the
SAMPLE
key several times to display
dout
.
Step 3 Press
the
key.
Step 4 Press
the
SAMPLE
key three times to display
data 0
.
Step 5 Press
the
RE-ZERO
key to display
data 2
.
Step 6 Press
the
key to store the setting.
Step 7 Press
the
CAL
key to return to the weighing mode.