12-5. Data Memory for Comparator Settings
The data memory function can store 20 sets of upper and lower limit values for the
comparator mode. The reference value or tolerance value for the comparator mode can not
be stored in memory.
The upper and lower limit values in memory can be recalled easily using the
MODE
key
and used for weighing.
The upper and lower limit values in memory can be recalled and changed.
Note
The recalled upper and lower limit values can be changed, using the digital input
mode or the weighing input mode. The digital input mode enters the upper and lower
limit values using the keys. The weighing input mode uses a sample to store the
upper and lower limit values.
12-5-1. Storing the upper and lower limit values
To store new upper and lower limit values: Recall the stored upper and lower limit values to
be changed ("
C01
" to "
C20
") . Then, change the recalled upper and lower limit values using
the digital input mode or the weighing input mode, and store the new value.
Note
While the data memory function is in use, unit selection using the
MODE
key is
not available.
If
"
Clr
" appears blinking in the upper left of the display, delete the data in memory.
Refer to "12-1. Notes on Using Data Memory".
Step 1 Press the
MODE
key to select a unit to be
used for storage.
Step 2 Set the "Data memory (
data
)" parameter to "
4
".
Step 3 Press and hold the
key until the
balance enters the upper and lower limit
values confirmation mode. The upper limit
value last selected is displayed.
Step 4 Select the comparator number to be used, using the following keys.
RE-ZERO
key..... To increase the comparator number by one.
MODE
key ......... To decrease the comparator number by one.
Each time the
RE-ZERO
key or
MODE
key is pressed, the displayed value
changes as follows:
↔
C03
HI
↔
C03
LO
↔
C04
HI
↔
C04
LO
↔
Step 5
To change the selected upper and lower limit values:
To use the digital input mode, press the
SAMPLE
key to enter the digital input mode.
Go to "Digital input mode" on the next page.
To use the weighing input mode, press the
SAMPLE
key, then press and hold the
MODE
key to enter the weighing input mode. Go to "Weighing input mode" on the
next page.
No.3. Comparator number