65
Notes
Using the "
CN:mm
" command, the comparator limit values can be recalled.
The upper limit value recalled can be output using the "
?HI
" command.
The lower limit value recalled can be output using the "
?LO
" command.
The upper limit value can be changed using the "
HI:
" command.
The lower limit value can be changed using the "
LO:
" command.
"
mm
" indicates a two-digit numerical value 01 to 20, which corresponds to
C01
-
C20
.
Digital input mode
In the digital input mode, the upper and lower limit values are
entered digitally using the keys. Use the following keys to store
upper and lower limit values in digital input mode.
SAMPLE
key ... To select the digit to be changed.
RE-ZERO
key.. To change the value of the selected digit.
MODE
key ...... To switch the polarity.
key....... To store the upper and lower limit values.
Go to step 3 of the previous page.
CAL
key.......... To return to the upper and lower limit
values confirmation mode
Go to step 3 of the previous page.
MODE
key
Press and hold to go to the weighing input mode.
Weighing input mode
In the weighing input mode, a sample is placed on the pan to store the upper and lower limit
values.
Notes
Pressing
the
CAL
key will interrupt the operation and the
balance will return to the upper and lower limit values
confirmation mode (step 3 in "Storing the upper and lower
limit values").
To go to the digital input mode, press and hold the
MODE
key.
Step 1 The first display in the weighing input mode depends on the
comparator number selected in step 4 in "Storing the upper and
lower limit values".
For example, when "
C03 H1
" is selected in step 4, the display is
the current weight value and the comparator number with
HI
blinking.
Step 2 Place a container on the weighing pan, if necessary.
Press
the
RE-ZERO
key to set the display to zero.
Step 3 Place a sample corresponding to the upper limit value,
on the pan or in the container.
Step 4 Press the
key to store the upper limit value.
Value Digit
Polarity
Store
C03
HI
Upper limit value
Comparator
number
Upper
limit
Store
Place a sample
corresponding to the
upper limit value
Summary of Contents for GF-30K
Page 7: ...5...
Page 95: ...93 20 1 External Dimensions Unit mm...
Page 102: ...100 MEMO...
Page 103: ......