67
12-6. Data Memory for Tare Value
The data memory function can store 20 sets of tare values for weighing.
The tare value in memory can be recalled easily using the
MODE
key and used for weighing.
The tare value in memory can be recalled and changed.
Notes
The recalled tare value can be changed, using the digital input mode or the weighing
input mode. The digital input mode enters the tare value using the keys. The
weighing input mode uses a sample tare container to store the tare value.
The NET indicator illuminates during tare operation.
12-6-1. Storing the tare value
To store a new tare value: Recall the stored tare value to be changed ("
t01
" to "
t20
"). Then,
change the recalled tare value using the digital input mode or the weighing input mode, and
store the new value.
Notes
The recalled tare value can be changed, using the digital input mode or the
weighing input mode. The digital input mode enters the tare value using the keys.
The weighing input mode uses a sample tare container to store the tare value.
The NET indicator illuminates during tare operation.
When
the
RE-ZERO
key is pressed with nothing placed on the weighing pan, zero
is displayed, The NET indicator does not illuminate.
"
t--
" appears when a tare operation is performed without using the tare value
stored in memory.
While the data memory function is in use, unit selection using the
MODE
key is
not available.
If
"
Clr
" appears blinking in the upper left of the display, delete the data in memory.
Refer to "12-1. Notes on Using Data Memory".
Step 1 Press the
MODE
key to select a unit to be used for storage.
Step 2 Set the "Data memory (
data
)" parameter to "
5
".
Step 3 Press and hold the
key until the balance
enters the tare value confirmation mode. The tare
value last selected is displayed.
Step 4 Select the tare number to be used, using the following keys.
RE-ZERO
key..... To increase the tare number by one.
MODE
key ......... To decrease the tare number by one.
CAL
key ............ To cancel it and to return to the weighing mode.
Each time the
RE-ZERO
key or
MODE
key is pressed, the displayed value
changes as follows:
↔
t03
↔
t04
↔
........
↔
t20
↔
t01
↔
Step 5 To change the selected tare value:
To use the weighing input mode, press the
SAMPLE
key to enter the weighing input
mode. Go to "Weighing input mode" on the next page.
To use the digital input mode, press the
SAMPLE
key, then press and hold the
MODE
key to enter the digital input mode. Go to "Digital input mode" on the next page.
No.3, Tare number