92
GF-8K GF-8K2 GF-10K GF-12K GF-20K GF-30K GF-32K
Weighing capacity
8.1 kg
10.1 kg
12 kg
21 kg
31 kg
31 kg
8.1008 kg
31.008 kg
Maximum display
8.10084 kg
2.10009 kg *1
10.10084 kg 12.0084 kg 21.0084 kg 31.0084 kg
6.1009 kg *1
Minimum weighing value (1 digit) 0.01 g 0.1 g / 0.01 g
0.01 g
0.1 g
1 g / 0.1 g
Repeatability
(Standard deviation)
0.01 g
0.05 g / 0.01 g
0.01 g
0.1 g
0.5 g / 0.1 g
Linearity
±0.03 g ±0.1 g/ ±0.02g
±0.03 g
±0.2 g
±1 g / ±0.2 g
Stabilization time
(Typical
at
FAST
)
Approx. 1.5 seconds
Sensitivity drift,
(10°C ~ 30°C / 50
°
F ~ 86
°
F) ±2 ppm/°C ±3 ppm/°C ±2 ppm/°C
±3 ppm/°C
±5 ppm/°C
Operating environment
5°C to 40°C (41°F to 104°F), 85%RH or less (No condensation)
Internal mass
not available
Weighing data of data
memory
40 data
Time and clock function
not available
Display refresh rate
5 times/second or 10 times/second
Minimum unit
mass
0.01 g
0.1 g
0.01 g
0.1 g
1g
Coun
tin
g
mode Number of
samples
10, 25, 50 or 100 pieces
Minimum 100%
reference mass
1 g
10 g
1 g
10 g
100g
Percent mode
Minimum 100%
display
0.01 %, 0.1 %, 1 % (Depends on the reference mass stored.)
Interface
(Provided as standard)
RS-232C with Windows Communication Tools Softare WinCT
External calibration weight
2 kg, 3 kg, 4 kg, 5 kg,
6 kg, 7 kg, 8 kg
2 kg, 3 kg,
4 kg, 5 kg,
6 kg, 7 kg,
8 kg, 9 kg,
10 kg
5 kg, 10 kg 10 kg, 20 kg
20 kg, 30 kg
Weighing pan
270 x 210mm
External dimensions
300(W) x 355(D) x 111(H) mm
Power supply &
AC adapter type
Power consumption: Approx. 11VA (supplied to the AC adapter )
Confirm that the adapter type is correct for the local voltage and power receptacle type.
Weight
Approx. 8.3kg
Dust and water protection
Complying with IP65
*1: The balance allows weighing using a precision range, even with a heavy tare placed on the pan. (Smart range function)
Summary of Contents for GF-30K
Page 7: ...5...
Page 95: ...93 20 1 External Dimensions Unit mm...
Page 102: ...100 MEMO...
Page 103: ......