GX-A/GF-A Maintenance manual
Page 35
Step 4. In the D2 display, place an external mass on the center of the pan and at four
positions half the distance from the center of the pan to the edge. Adjust the corner
load adjusting screws so that the difference between the values in the center and
each position will be within the specification. (If a condition in the illustration below
occurs, turn the corner load adjusting screws as indicated by the arrows.)
Step 5. When adjusting the corner load, a drift of the zero point may be generated. After
adjusting, load the mass listed in table below on the front, back, left and right of the
weighing pan, and work again after the drift of the center zero point is confirmed to
be within ±3 digits.
Model (GX-A/GF-A series)
Masses
Specifications
Drift value of
Zero point
123 (Only GF-A)
100g
±
0.006g
±
0.003g
203
303
200g
±
0.008g
403
±
0.007g
603
300g
±
0.007g
1003
500g
±
0.007g
1603
±
0.008g
1202 (Only GF-A)
1kg
±
0.06g
±
0.03g
2002
3002
2kg
±
0.08g
4002
±
0.07g
6002
3kg
±
0.07g
10002
4kg
±
0.10g
6001
3kg
±
0.3g
±
0.3g
10001
4kg
If the corner load error at the front of the weighing pan is plus counts, turn the four
corner adjust screws counterclockwise.
If the corner load error at the front of the weighing pan is minus counts, turn the four
corner adjust screws clockwise.
◎
◎
◎
◎
◎
◎
◎
◎
Weighing pan
Four corner adjust screw
Center
Summary of Contents for GF-A Series
Page 51: ......