GX-A/GF-A Maintenance manual
Page 45
Model (GX-A series)
Standard masses
Specifications
203
200g
±
0.010g
303
300g
403
400g
603
600g
1003
1000g
1603
2002
2kg
±
0.10g
3002
3kg
±
0.15g
4002
4kg
6002
5kg
10002
6001
5kg
±
0.5g
10001
(6) Calibrate using the internal mass and verify the fine adjustment error (operation (4)-(5)).
Perform this three times, verify that the all data be within the specifications.
4
. check the operation of the electronically controlled load
①
W
ith D2 MODE, when loaded with electronically controlled load , the value and
stability of the span in the table below have to be checked.
Confirm that it is within
range.
W
ith the key operation (pressing the SAMPLE key while pressing the MODE key),
you can load the electronically controlled load .
Also it can be operated in the following command.
FP7:0
electronically controlled load ON
FP7:1
electronically controlled load OFF
※
When the electronically controlled load is ON, the mark lights on the left
side of the upper part of the display.
②
Electronically controlled load,confirm that the repeatability [standard deviation SD] (10
times or 5 times) of the span at the time of load is within the range of Table below.
Check mode with VE TEST, it is possible to repeat load of electronically controlled load
and effectiveness of repeatability calculation.
Model (GX-A series)
D2span
D2 stability
VE TEST
Span
reproducibility σ
104,204,304
1.5~3.5
g
Within 0.1 mg / 5 seconds
Within 0.1 mg
123,203,303,403,603,1003,1603
3~12
g
Within
0.001mg
/ 5 seconds
Within 0.001g
1202,2002,3002,4002,6002,10002
15~55
g
Within
0.01mg
/ 5 seconds
Within 0.01g
6001,10001
15~55
g
Within
0.1mg
/ 5 seconds
Within 0.1g