53
Example1
(
Comparison when the weighing data is stable or overloaded, excluding "near
zero" ± 20 digit, upper limit and lower limit.
)
Selecting a comparator mode
Step1 Press and hold the SAMPLE key until
ba5fnc
of the function table is displayed, then
release the key.
Step2 Press the SAMPLE key several times to display
Cp fnc
.
Step3 Press the PRINT key.
Step4 Press the RE-ZERO key several times to display
( "
" always compare).
.
Step5 Press the SAMPLE key several times to display
.
Step6 Press the RE-ZERO key several times to display
( " " ± 20 digit is not
compared.)
Step7 Press the SAMPLE key several times to move to
Cpin
.
Step8 Press the RE-ZERO key several times to display
. ( " " reference value is set.
digital input)
Step9 Press the PRINT key to store the selected mode.
Entering the upper and lower limit values
Step10 With displayed, press the PRINT key to display .
Step11 Press the PRINT key. The current setting of the upper limit value is displayed with all of the
digits blinking. When the current setting is not to be changed, press the PRINT or CAL
key to go to step 12.
When the current setting is to be changed, press the RE-ZERO key and store the following
keys.
SAMPLE key
To select the digit to change the value.
RE-ZERO key
To change the value of the digit selected.
MODE key
To switch the polarity.
PRINT key
To store the new setting and go to step10.
CAL key
To cancel the new setting and go to step 10.
Step12
When
is displayed, pressing the PRINT key will display the currently set value.
If changing the setting value, it can be registered the tolerance value with the following keys.
For tolerance value, enter the value with the reference value set to 100%.
SAMPLE key
Move the blinking digit.
RE-ZERO (+)key
Change the value of the blinking digit.
MODE (-)key
Change the value of the blinking digit.
PRINT key
Register and go to Step13
CAL key
Cancel and go to Step13
Step13
Press the CAL key to return to the weighing display.