5-stage comparison result
5-stage comparison - display, contact output
Weighing value
Result
Display
Contact output
Buzzer
Threshold value Formula
HH HI OK LO LL
SUL
UL
LL
SLL
SUL < WV
HH HI blinking
Short
UL < WV
≤
SUL HI HI illuminated
Short
Open
LL
≤
WV
≤
UL
OK OK illuminated
Short
SLL
≤
WV < LL
LO LO illuminated
Open
Short
WV < SLL LL LO blinking
Short
Weighing value
:
WV, Second upper limit
:
SUL, Upper limit
:
UL, Lower limit
:
LL, Second lower limit
:
SLL
6.3.
Comparison Method Selection (Tolerance Range and Center
Value or Threshold Values)
Setting Example
Comparison always ......................................................................
5-stage comparison ......................................................................
Comparison excluding ±5 digits (counts) from zero ...................
Comparison of both polarities ......................................................
Factory setting
Center value using numerical input ...................................................
or
Center value using sample weight ....................................................
Procedure
Step 1 Press and hold the SAMPLE key to enter the function table.
is displayed.
Step 2 Press the SAMPLE key to display
(Comparator).
Step 3 Press the PRINT key to enter
(Comparator).
Step 4 Press the RE-ZERO key several times to display
(Comparison always).
Step 5 Press the SAMPLE key to display
(Comparison
method).
Step 6 Press the RE-ZERO key to display
(5-stage
comparison).
Step 7 Press the SAMPLE key to display
(Near zero).
Step 8 Press the RE-ZERO key several times to display
(Comparison excluding ±5 digits (counts) from zero).
Step 9 Press the SAMPLE key several times to display
(Input method).
Step 10 Press the RE-ZERO key several times to select input method.
(Center value using numerical input) or,
(Center value using sample weight)
Step 11 Press the PRINT key to store new settings and use them.
or