25
With the “RsKey” test mode function, a personal computer can be used as an external weight
indicator for the balance. (To do this, set the balance data output mode to stream mode.)
4-9 Commands
Command list
Commands to query weighing data
C
Cancels the
S
or
SIR
command.
Q
Requests the weighing data immediately.
S
Requests the weighing data when stabilized.
SI
Requests the weighing data immediately.
SIR
Requests the weighing data continuously.
E
SC
P
Requests the weighing data when stabilized.
Commands to control the balance
CAL
Same as the CAL key.
OFF
Turns the display off.
ON
Turns the display on.
P
Same as the ON:OFF key
PRT
Same as the PRINT key
R
Same as the RE-ZERO key
SMP
Same as the SAMPLE key.
U
Same as the MODE key
T
Same as the RE-ZERO key
Z
Same as the RE-ZERO key
E
SC
T
Same as the RE-ZERO key
?ID
Requests the ID number.
?SN
Requests the serial number.
?TN
Requests the model name.
Note
E
SC
: 1Bh in ASCII code
Commands to query memory data
MCL
Deletes all data in memory.
MD:nnn
Deletes weighing data with the data number nnn.
?MA
Outputs all data in memory.
?MQnnn
Outputs weighing data with the data number nnn.
?MX
Outputs the number of weighing data in memory.
Note