38
Pcs On the
ACAI
Weighing Pan
Addition Range
60
63~122
10
13~26
70
73~138
20
23~49
80
83~152
30
33~70
90
93~166
40
43~89
100
103~299
50
53~106
over 200
203~492
As you add, the ACAI
annunciator will be
ON as long as you
are in range.
When you stop adding
and the display
becomes
STABLE, the ACAI
annunciator will blink.
After the new unit
weight is calculated,
the annunciator will
disappear.
3. Continue adding pieces within the ACAI range until you have reached a sample size
as large as the largest number of pieces that you will be counting.
When you have added the maximum number of pieces required, remove the
sample pieces and start your counting job.
10-3. ACAI Manual Operation
The ACAI procedure can also be controlled manually. The ACAI will not recalculate
the unit weight until the
ENTER
key is pressed (as long as it is at the proper time
and the guidelines in the ACAI Notes section have been followed).
The ACAI manual mode is controlled by F-Function
f-02-1
, set at “
2
”.
To start ACAI manual operation, the unit weight must be registered and the samples
still on the weighing pan.
1. Add pieces within the nearest ACAI range
(refer to table in the previous section). The
ACAI annunciator will stay ON as long as you
are within ACAI range.
2. Wait until the display becomes stable and
press the
ENTER
key.
When the new unit weight is calculated, the
indicator will blink for a moment and then turn off.
3. Continue adding pieces within the ACAI range until you have reached a sample size
as large as the largest number of pieces that you will be counting.
When you have added the maximum number of pieces required, remove the
sample pieces and start your counting job.
STABLE
ACAI
STABLE
|
ACAI
STABLE
|
ACAI
888888
21
COUNT
WEIGHT
UNIT WEIGHT
ACAI
888
2.331
8
8
111.000