44
13-3. Data Output Mode
The Data Output Modes and Parameters are set by F-Functions in F-06-X as
described in the “9-2. F-Functions”
To control the scale using commands from an external device, see “13-6. Command
Mode”.
Refer to “13-6. Command Mode” about the output data format.
Data Output Mode (
f-06-01
)
Key Mode (
f-06-01
="
0
")
When the weight display is stable, data is sent by pressing the
key. The
count display will blink when the data has been sent.
Stream Mode (
f-06-01
="
1
")
Data is sent continuously. The data-update rate is approximately 10 times per
second for
f-06-03
="
2
". For
f-06-3
="
0
" or "
1
", the interval between continuous
data is approximately 2 seconds.
Auto-print Mode A (
f-06-01
="
2
")
Data is sent if the weight display is stable at +5d (weighing display division) and
above. The next transmission can not occur until after the weight display falls below
+5d.
Auto-print Mode B (
f-06-01
="
3
")
Data is sent if the weight display is stable at ±5d (weighing display division) and
above/below. The next transmission can not occur until after the weight display falls
between –5d and +5d.
To use with the UFC format, refer to "13-7. Using UFC (Universal Flex Coms)
Function"
Data to be Sent (
f-06-02
)
Select which data is to be sent by keying in a
0
or
1
for the data: ID No., PCS
(count), weight or unit weight.
Example:
Key in
1
1
0
0
to
display
1100
, this setting would
send only the ID number and the
count.
Data Format (
f-06-03
)
Format for AD-8121 MODE 1 or 2. (
f-06-03
="
0
")
Format for AD-8121 MODE 3. (
f-06-03
="
1
")
Format for general apparatuses, computers, etc. (
f-06-03
="
2
")
‘
0
’ not to send data
‘
1
’ to send data
8
f-06-02
88Clear8
8
ai
1100
0
00
ID No.
PCS (count)
Weight
Unit Weight