background image

 

3

PREPARATION 

Installing Batteries / Connecting The AC adapter

 

Installing Batteries 

CAUTION 

When replacing batteries, be careful 
not to apply too much force to the 
weighing pan. An excessive force 
could damage the scale. 

Remove the battery cover and insert six 
batteries (R6P / LR6 / AA size) into the 
battery compartment, taking extreme 
care to that the polarities plus and minus 
are observed. Be sure to lock the 
battery cover again after installing the 
batteries. 

 
 
 

AC adapter (optional) 

CAUTION 

When using an AC adapter, the scale 
does not comply with the IP65. 

 
The AC input requirement could be 100, 
120, 220, 230 or 240 volts (50/60Hz) 
depending on the area where used, so 
please verify that the adapter is correct. 

Plug the AC adapter to the AC adapter 
socket inside the battery compartment. 

Attach the battery cover and fit the 
rubber cap. 

 

Setting up the HL-WP

 

CAUTION 

Avoid installing the scale in direct sunlight, as that may cause discoloration 
or a malfunction. Place your HL-WP on a firm weighing table so that the 
scale is level. (The scale will not perform accurately when it is not level.) 

 

Place the scale on the firm surface and adjust the feet so that the sprit level 
shows the scale is level for an accurate weighing. 

New six batteries (R6P/ LR6/ AA size)

Optional AC Adapter 

AC Adapter jack

 

Close the rubber cap 

Battery cover lock

Summary of Contents for HL-WP Series

Page 1: ...HL 1000WP HL 300WP HL 3000WP HL 3000LWP 1WMPD4000754B Digital Scale HL WP Series This document hosted by www oldwillknottscales com...

Page 2: ...AC adapter for the scale Other adapters may cause damage The AC adapter is optional Internal service or adjustments to this product should be performed by a qualified person CAUTION Avoid installing...

Page 3: ...tery cover Battery compartment AC adapter jack Batteries are not included Rubber cap Battery cover lock U S A models only 04 76 3k oz 04l12 3k oz 0 1 000k lb 01200 0kg STABLE indicator Indicates when...

Page 4: ...put requirement could be 100 120 220 230 or 240 volts 50 60Hz depending on the area where used so please verify that the adapter is correct Plug the AC adapter to the AC adapter socket inside the batt...

Page 5: ...ress the ON OFF key to turn on the scale Then the scale displays all segments for a few seconds and then 0 will be displayed 2 Start weighing If you do not use a container for weighing Verify the read...

Page 6: ...2g HL 3000WP 3000g 0 2g HL 3000LWP 3000g 0 2g 1 Enter calibration mode Press the ON OFF key to turn the power off While pressing both the RE ZERO key and the UNITS key press the ON OFF key Cal will be...

Page 7: ...value 1 Enter calibration mode Press the ON OFF key to turn the power off While pressing both the RE ZERO key and the UNITS key press the ON OFF key Cal will be displayed 2 Select a digit to be change...

Page 8: ...lb 950 g 2 1 lb Calibration weight 1000g 0 1g 300g 0 02g 3000g 0 2g Accessories This manual Options AC adapter The Value of Gravity at Various Locations Amsterdam 9 813 m s2 Manila 9 784 m s2 Athens...

Page 9: ...8 World Map...