WEIGHING 

Before weighing 

Automatic Power-Off Function

 

If the scale is left ON and the stable indicator is displayed, the automatic 
power-off function turns the power off after approximately 5 minutes 
(factory setting). To toggle the function disabled/enabled, press the 
[ON/OFF] key while pressing the [RE-ZERO] key. The scale stores the 
setting and it will be shown by pressing both the [RE-ZERO] and 
[UNITS] keys in the weighing mode. (See “CALIBRATION” about the 
[UNITS] key.) 

 

 

poff0

e

 

Function disabled.

       

 

 

poff1

e

 

Function enabled.

  

Error message

 

 

ee

e

ee

 

   Overload 

 ----- 

    The scale zero is out of range 

 

ea

lb

a

 

   Low battery 

Calibration may be required before weighing

 

Read “CALIBRATION” first and if necessary, calibrate your scale for 
accurate weighing. 

 

Weighing procedures

 

1. Turn on the scale.

 

Press the [ON/OFF] key to turn on the scale. Then the scale displays all 
segments for a few seconds and then 

   0 

  will be displayed. 

 

2. Start weighing

 

If you do not use a container for weighing, 

Verify the reading is 

   0 

. If not, press the [RE-ZERO] key to display 

   0 

. Place objects on the weighing platform to weigh. When the reading 

becomes stable, the stable indicator is displayed. 

If you use a container for weighing,

 

Place an empty container on the platform.   
Wait for the stability indicator to be displayed and press the [RE-ZERO] 
key. Place the objects to be weighed in the container. 

Select the weighing unit, --- Only for U.S.A. models --- 

If you use the U.S.A. model, press the [UNITS] key to select weighing unit.   
Normally scales power up in “lb” or “oz” when shipped from the factory. 

To 

power up in other units

, turn the scale off. Press and hold the [UNITS] key, 

then press the [ON/OFF].key The scale will start with the next unit.   
Note) HL-300WP: oz

Æ

g

Æ

oz

Æ

 

 HL-1000WP/3000WP/3000LWP: 

lb

Æ

oz

Æ

lb-oz

Æ

g

Æ

lb

Æ

 

Summary of Contents for HL-WP Series

Page 1: ...HL 1000WP HL 300WP HL 3000WP HL 3000LWP 1WMPD4000754B Digital Scale HL WP Series This document hosted by www oldwillknottscales com...

Page 2: ...AC adapter for the scale Other adapters may cause damage The AC adapter is optional Internal service or adjustments to this product should be performed by a qualified person CAUTION Avoid installing...

Page 3: ...tery cover Battery compartment AC adapter jack Batteries are not included Rubber cap Battery cover lock U S A models only 04 76 3k oz 04l12 3k oz 0 1 000k lb 01200 0kg STABLE indicator Indicates when...

Page 4: ...put requirement could be 100 120 220 230 or 240 volts 50 60Hz depending on the area where used so please verify that the adapter is correct Plug the AC adapter to the AC adapter socket inside the batt...

Page 5: ...ress the ON OFF key to turn on the scale Then the scale displays all segments for a few seconds and then 0 will be displayed 2 Start weighing If you do not use a container for weighing Verify the read...

Page 6: ...2g HL 3000WP 3000g 0 2g HL 3000LWP 3000g 0 2g 1 Enter calibration mode Press the ON OFF key to turn the power off While pressing both the RE ZERO key and the UNITS key press the ON OFF key Cal will be...

Page 7: ...value 1 Enter calibration mode Press the ON OFF key to turn the power off While pressing both the RE ZERO key and the UNITS key press the ON OFF key Cal will be displayed 2 Select a digit to be change...

Page 8: ...lb 950 g 2 1 lb Calibration weight 1000g 0 1g 300g 0 02g 3000g 0 2g Accessories This manual Options AC adapter The Value of Gravity at Various Locations Amsterdam 9 813 m s2 Manila 9 784 m s2 Athens...

Page 9: ...8 World Map...

Reviews: