CALIBRATION 

When to calibrate

 

Calibration may be required when the scale initially installed, if it is 
moved a substantial distance, or in accordance with local regulations. 
This is necessary because the weight of a mass in one location is not 
necessarily the same in another location. Also, with time and use, 
mechanical deviations can occur. 
 

The position of the [UNITS] key for the Calibration

 

The “g” model has no 
[UNITS] key, but the key 
used to enter calibration is 
located under the overlay. 

 

 

Calibration using a calibration weight

 

Calibrate the scale with calibration weights at full capacity. 

HL-1000WP : 

1000g ± 0.1g 

HL-300WP: 

300g ± 0.02g 

HL-3000WP : 

3000g ± 0.2g 

HL-3000LWP : 

3000g ± 0.2g   

 

1. Enter calibration mode. 

Press the [ON/OFF] key to turn the power off. 
While pressing both the [RE-ZERO] key and the [UNITS] key, press the 
[ON/OFF] key. 

 Cal  

 will be displayed.

 

 

2. Zero calibration 

Press the [RE-ZERO] key to calibrate zero. 

 Cal 0 

  will be displayed. 

Wait for the stable indicator to be displayed and press the [RE-ZERO] key. 

 Cal

a

  will be displayed in a few seconds. If only the zero calibration is 

to be done, press the [UNITS] key. The scale will display 

 end 

 and 

automatically return to the weighing mode.

 

 

3. Span calibration 

Place the weight at the center of the platform for accurate calibration.   Wait 
for the stable indicator to be displayed and then press the [RE-ZERO] key. 
The display will show 

 end 

 and the scale will automatically return to the 

weighing mode.

 

 
 
 
 

Location of the [UNITS] key 

Summary of Contents for HL-WP Series

Page 1: ...HL 1000WP HL 300WP HL 3000WP HL 3000LWP 1WMPD4000754B Digital Scale HL WP Series This document hosted by www oldwillknottscales com...

Page 2: ...AC adapter for the scale Other adapters may cause damage The AC adapter is optional Internal service or adjustments to this product should be performed by a qualified person CAUTION Avoid installing...

Page 3: ...tery cover Battery compartment AC adapter jack Batteries are not included Rubber cap Battery cover lock U S A models only 04 76 3k oz 04l12 3k oz 0 1 000k lb 01200 0kg STABLE indicator Indicates when...

Page 4: ...put requirement could be 100 120 220 230 or 240 volts 50 60Hz depending on the area where used so please verify that the adapter is correct Plug the AC adapter to the AC adapter socket inside the batt...

Page 5: ...ress the ON OFF key to turn on the scale Then the scale displays all segments for a few seconds and then 0 will be displayed 2 Start weighing If you do not use a container for weighing Verify the read...

Page 6: ...2g HL 3000WP 3000g 0 2g HL 3000LWP 3000g 0 2g 1 Enter calibration mode Press the ON OFF key to turn the power off While pressing both the RE ZERO key and the UNITS key press the ON OFF key Cal will be...

Page 7: ...value 1 Enter calibration mode Press the ON OFF key to turn the power off While pressing both the RE ZERO key and the UNITS key press the ON OFF key Cal will be displayed 2 Select a digit to be change...

Page 8: ...lb 950 g 2 1 lb Calibration weight 1000g 0 1g 300g 0 02g 3000g 0 2g Accessories This manual Options AC adapter The Value of Gravity at Various Locations Amsterdam 9 813 m s2 Manila 9 784 m s2 Athens...

Page 9: ...8 World Map...

Reviews: