Gravity compensation

 

 

If you have calibration weights at the capacity of the scale, you can calibrate 
the scale with the weights instead of compensating for acceleration due to 
gravity. 

 

The scale was calibrated at 9.798 m/sec

2

 , at the factory. If the acceleration due 

to gravity at your location is not same, compensate the scale by adjusting the 
acceleration due to gravity. (Refer “The Value of Gravity at Various Locations” 
and “World Map” and find your location value.) 

 

1. Enter calibration mode. 

Press the [ON/OFF] key to turn the power off. While pressing both the 
[RE-ZERO] key and the [UNITS] key, press the [ON/OFF] key. 

 Cal 

 will 

be displayed. 
 

2. Select a digit to be changed. 

Press the [UNITS] key. The previously set value, for example

 

 979,8 

, will 

be displayed. A digit after the comma “,” will be changed by pressing the 
[RE-ZERO] key. Press the [UNITS] key to move the comma “,” and select 
a digit to be changed. 

 

3. Set the new value 

Change the value of the gravity acceleration using the [RE-ZERO] and 
[UNITS] keys.   

 

4. Save the value in memory. 

Press and hold the [UNITS] key until 

 end 

 is displayed. The scale will 

automatically return to the weighing mode.

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Summary of Contents for HL-WP Series

Page 1: ...HL 1000WP HL 300WP HL 3000WP HL 3000LWP 1WMPD4000754B Digital Scale HL WP Series This document hosted by www oldwillknottscales com...

Page 2: ...AC adapter for the scale Other adapters may cause damage The AC adapter is optional Internal service or adjustments to this product should be performed by a qualified person CAUTION Avoid installing...

Page 3: ...tery cover Battery compartment AC adapter jack Batteries are not included Rubber cap Battery cover lock U S A models only 04 76 3k oz 04l12 3k oz 0 1 000k lb 01200 0kg STABLE indicator Indicates when...

Page 4: ...put requirement could be 100 120 220 230 or 240 volts 50 60Hz depending on the area where used so please verify that the adapter is correct Plug the AC adapter to the AC adapter socket inside the batt...

Page 5: ...ress the ON OFF key to turn on the scale Then the scale displays all segments for a few seconds and then 0 will be displayed 2 Start weighing If you do not use a container for weighing Verify the read...

Page 6: ...2g HL 3000WP 3000g 0 2g HL 3000LWP 3000g 0 2g 1 Enter calibration mode Press the ON OFF key to turn the power off While pressing both the RE ZERO key and the UNITS key press the ON OFF key Cal will be...

Page 7: ...value 1 Enter calibration mode Press the ON OFF key to turn the power off While pressing both the RE ZERO key and the UNITS key press the ON OFF key Cal will be displayed 2 Select a digit to be change...

Page 8: ...lb 950 g 2 1 lb Calibration weight 1000g 0 1g 300g 0 02g 3000g 0 2g Accessories This manual Options AC adapter The Value of Gravity at Various Locations Amsterdam 9 813 m s2 Manila 9 784 m s2 Athens...

Page 9: ...8 World Map...

Reviews: