18
6.4.
Percent Mode (%)
The percent mode displays the weight value in percentage compared to a 100% reference
mass and is used for target weighing or checking the sample variance.
Selecting the percent mode
1 Press
the
MODE
key to select the unit
%
(Percent mode).
If the percent mode can not be selected, refer to "5. Weighing
Units".
Storing the 100% reference mass
2 Press
the
RANGE
key to enter the 100% reference mass
storing mode.
Even in the storing mode, pressing the
MODE
key will
switch to the next mode.
3 Place a container on the weighing pan, if necessary. Press
the
RE-ZERO
key to cancel the weight (tare). The balance
displays
100 0
%
.
4 Place the sample to be set as the 100% reference mass on
the pan or in the container.
5 Press the
key to store the reference mass. The
balance displays
100.00
%
. (The decimal point position
depends on the reference value. The reference mass is
stored in non-volatile memory, and is maintained even if the
AC adapter is removed.)
Note
The decimal point position depends on the 100% reference mass.
100% mass
Minimum display
0.0100g to 0.0999g
1%
0.1000g to 0.9999g
0.1%
1.0000g to weighing
capacity
0.01%
If the balance judges that the mass of the sample is too light
(under 0.01g) to be used as a reference, it displays
lo
.
A 100% reference mass can be stored in the non-volatile
memory and is maintained even if the AC adapter is
removed.
6 Remove
the
sample.
Reading the percentage
7 Place a sample to be compared to the reference mass on
the pan. The displayed percentage is based on the 100%
reference mass.
Pan
Container
100% mass
Place sample
Percentage
Summary of Contents for HR-202i
Page 1: ...Analytical Balance HR 300i HR 202i 1WMPD4000912C ...
Page 7: ...5 ...
Page 56: ...54 15 1 External Dimensions Unit mm ...
Page 62: ...60 MEMO ...