28
10.2.
Details of the Function Table
Class
Item and Parameter
Description
0
Fast response, sensitive value
FAST
1
MID.
Cond
Condition
2
Slow response, stable value
SLOW
Can be changed by response
adjustment.
0
Stable range is ±1 digit
1
#1
5t-b
Stability band width
2
Stable range is ±3 digits
The stabilization indicator illuminates
when the display fluctuation is within
the range per second.
0
OFF
1
Normal
2
Strong
trc
Zero tracking
3
Very strong
Keeps zero display by tracking zero
drift.
0
5 times/second
5pd
Display refresh rate
1
10 times/second
Period to refresh the display
0
Point (.)
pnt
Decimal point
1
Comma (,)
Decimal point format
0
OFF
ba5fnc
Environment
Display
p-on
Auto display-ON
1
ON
Turns on the weighing mode display
when the AC adapter is connected.
0
Key mode
Accepts the PRINT key only when
the display is stable.
1
Auto print mode A
(Reference = zero)
2
Auto print mode B
(Reference = last stable value)
Outputs data when the display is
stable and conditions of
ap-p
,
ap-b
and the reference value are met.
prt
Data output mode
3
Stream mode
Outputs data continuously.
0
Plus only
Displayed value>Reference
1
Minus only
Displayed value<Reference
ap-p
Auto print polarity
2
Both
Regardless of displayed value
0
10 digits
1
100 digits
ap-b
Auto print difference
2
1000 digits
Difference between reference value
and displayed value
#2
0
No output
5-id
ID number output
1
Output
Selects whether or not the ID
number is output.
0
No pause
pU5e
Data output pause
1
Pause (1.6 seconds)
Selects the data output interval.
0
Not used
at-f
Auto feed
1
Used
Selects whether or not automatic
feed is performed.
0
No output
1
AD-8121 format
info
GLP output
2
General data format
Selects GLP output method.
For how to set time and date to be
added, refer to peripheral manual.
0
Not used
dout
Data output
ar-d
Zero after output
1
Used
Adjusts zero automatically after data
is output
: Factory settings.
#1 The unit of minimum display is digit. Example: If 1 mg display is selected using the
RANGE
key for the HR-300
i
, 1 mg is one digit.
#2 Usable minimum display of the balance is one digit. Example: In gram display, one digit is
0.1 mg for the HR-300
i
, and 0.01 mg for the HR-202