39
-----------
A
-
&
-
D
MODEL
----
HR-300i
S/N
-----
01234567
ID
------
LAB-0123
DATE
--
2004/09/30
START
TIME
----
12:34:56
WT
--
+123.4567
--
g
WT
--
+123.4612
--
g
WT
--
+123.4623
--
g
WT
--
+123.4634
--
g
WT
--
+123.4511
--
g
WT
--
+123.4534
--
g
WT
--
+123.4712
--
g
WT
--
+123.4645
--
g
END
TIME
----
12:40:15
SIGNATURE
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
Title block and end block
When mass values are recorded as GLP data, a "Title block" is inserted at the beginning
and an "End block" is inserted at the end of a group of mass values in the GLP report.
Notes
To output the report to an AD-8121B , use MODE 3 of the AD-8121B. If MODE1 is used,
select temporary dump print mode by pressing the
STAT.
key of the AD-8121B.
If the data memory function is used (except
data 0
), the "Title block" and "End block" can not
be output.
Key operation
1 With the weighing data displayed, press and hold the
key, then release the key.
5tart
is displayed. The "Title block" is output.
2 The weighing data is output according to the parameter setting of the data output mode
(
prt
) of the function table.
3 Press and hold the
key until
recend
is displayed, then release the key.
The "End block" is output.
Setting of "
info 1
"
Setting of "
info 2
"
AD-8121 printer format
General format
Space, ASCII 20h
<TERM>
Terminator, CR , LF or CR
CR
Carriage return, ASCII 0Dh
LF
Line feed, ASCII 0Ah
Title block
Manufacturer
Model
Serial number
ID number
Date
Time
Weighing data
Time
Signature
End block
The space
to fill date in.
The space
to fill start time in.
The space
to fill end time in.