51
Display Error
code
Description
Unit mass error
The sample unit mass for the counting mode is too light.
Storing and using it for counting will cause a counting error.
Add samples to reach the specified number and press the
key. Pressing the
key without adding samples
will shift the balance to the counting mode. But, for accurate
counting, be sure to add samples.
EC,E00
Communications error
A protocol error occurred in communications.
Confirm the format, baud rate and parity.
EC,E01
Undefined command error
An undefined command was received.
Confirm the command.
EC,E02
Not ready
A received command can not be processed.
Example:
The balance received a "
Q
" command, but not in the
weighing mode.
The balance received a "
Q
" command while processing a
RE-ZERO command. Adjust the delay time to transmit a
command.
EC,E03
Timeout error
If the timeout parameter is set to "
t-Up1
", the balance did not
receive the next character of a command within the time limit
of one second. Confirm the communication.
EC,E04
Excess characters error
The balance received excessive characters in a command.
Confirm the command.
EC,E06
Format error
A command includes incorrect data.
Example:
The data is numerically incorrect.
Confirm the command.
EC,E07
Parameter setting error
The received data exceeds the range that the balance can
accept. Confirm the parameter range of the command.
Other error code
If an error described above can not be released or other
errors are displayed, contact the local A&D dealer.
Summary of Contents for HR-202i
Page 1: ...Analytical Balance HR 300i HR 202i 1WMPD4000912C ...
Page 7: ...5 ...
Page 56: ...54 15 1 External Dimensions Unit mm ...
Page 62: ...60 MEMO ...