24
9.2.
Initializing the Balance
This function returns the following parameters to factory settings.
Calibration
data
Function
table
The sample unit mass value (counting mode),
100% reference mass value (percent mode)
External calibration weight and target mass value
Function switch settings ("9.1. Permit or Inhibit")
Note Be sure to calibrate the balance after initialization.
1 Press
the
ON:OFF
key to turn off the display.
2 While pressing and holding the
key and the
RANGE
key, press the
ON:OFF
key to display
p5
.
3 Press
the
RANGE
key to display
Clr
.
4 Press
the
key.
To cancel this operation, press the
CAL
key.
5 Press
the
RE-ZERO
key to display
Clr go
.
6 Press
the
key to initialize the balance.
The balance will automatically return to the weighing mode.