11.2.
GLP Report
Set the following parameters to output the report.
Refer to "13.2. Connection to peripheral equipment", for connection to an AD-8121B.
To print the report, set the "GLP output (
info
)" parameter to "
1
" and use MODE 3 of the
AD-8121B. If MODE1 is used, select temporary dump print mode by pressing the
STAT.
key of the AD-8121B.
To output the report to a personal computer using the RS-232C interface, set the "GLP
output (
info
)" parameter to "
2
".
If the time and date are not correct, adjust the settings of the AD-8121B.
Calibration report using an external weight
Key operation
1 Press and hold the
CAL
key to display
Cal
out
and release the key.
2
Cal
0
is displayed.
3 When updating the calibration mass value, press the
RANGE
key and proceed to step 4.
When using the preset calibration mass value, proceed to step 5.
4 Specify calibration mass value using the following keys.
RANGE
key ........ The key to select the blinking figure
RE-ZERO
(+)key .. The key to increase the value of the blinking figure.
MODE
(-)key........ The key to decrease the value of the blinking figure.
key ......... The key to store the new value.
5 When pressing the
RE-ZERO
key, the zero point is measured and the weight value is
displayed for a few seconds.
6 Place the displayed mass on the pan. Press the
key to measure the mass and the
weight value is displayed for a few seconds.
7 If GLP output is used,
glp
is displayed and the calibration report is output.
8 The balance returns to the weighing mode after this calibration.
Setting of "
info 1
"
Setting of "
info 2
"
AD-8121 printer format
General format
Space, ASCII 20h
<TERM>
Terminator, CR , LF or CR
CR
Carriage return, ASCII 0Dh
LF
Line feed, ASCII 0Ah
Manufacturer
Model
Serial number
ID number
Date
Time
Calibration type
Calibration weight
Signature
-----------
A
-
&
-
D
MODEL
----
HR-300i
S/N
-----
01234567
ID
------
LAB-0123
DATE
--
2004/09/30
TIME
----
12:34:56
CALIBRATED(EXT.)
CAL.WEIGHT
----
+200.0000
--
g
SIGNATURE
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
The space
to fill date in.
The space
to fill time in.