14.3.
Checking the Balance Performance and Environment
The balance is a precision instrument. When the operating environment or the operating
method is inadequate, correct weighing can not be performed. Place a sample on the pan
and remove it, and repeat this several times. If the balance seems to have a problem with
repeatability or to perform improperly, check as described below. If improper performance
persists after checking, contact the local A&D dealer for repair.
Checking that the operating environment or weighing
method is proper
Operating environment
Is the weighing table solid enough?
Is the balance level?
Is the operating environment free from vibration and drafts?
Is there a strong electrical or magnetic noise source such as a motor near the balance?
Weighing method
Does the weighing pan rim touch anything? Is the weighing pan assembly installed
correctly?
Is
the
RE-ZERO
key pressed before placing a sample on the weighing pan?
Is the sample placed in the center of the weighing pan?
Is the fine range breeze break ring installed for weighing with a minimum display of 0.01 mg
for the HR-202
i
?
Has the balance been warmed up for one hour before weighing?
Sample and container
Has the sample absorbed or lost moisture due to the ambient conditions such as
temperature and humidity?
Has the temperature of the container been allowed to equalize to the ambient temperature?
Is the sample charged with static electricity?
Is the sample of magnetic material such as iron? There are cautions about weighing
magnetic materials.
Checking that the balance performs properly
Check the balance performance using an external weight. Be sure to place the weight in the
center of the weighing pan.
Check the balance repeatability, linearity and calibrated value using an external weight with
a known value.
14.4.
Asking for Repair
If the balance needs service or repair, contact the local A&D dealer.
The balance is a precision instrument. Use much care when handling the balance and
observe the following when transporting the balance.
Use the original packing material for transportation.
Remove the weighing pan, pan support, breeze break ring and dust plate from the main
unit.