Page 42
HVG-WP Series
14.2.
The Complete Calibration Procedure
14.2.1.
Gravity Acceleration Correction
Step 1 Turn on the display.
Open the rear cal cover of the display unit.
Press and hold the
CAL
key to enter the calibration
mode.
Then
Cal 0
is displayed.
Step 2 Press the
ZERO
key to enter the gravity acceleration
correction mode.
Step 3 Set your local gravity acceleration using the following
keys.
∧
key
Selecting the number of a figure.
<
key
Selecting a figure.
Step 4 Press the
ENTER
key to store the new value.
Step 5 Press the
CAL
key again. Then the scale returns to
the normal weighing mode.
14.2.2.
Preparation
Step 6 Confirm the environmental conditions as follows:
Maintain a constant temperature and stable power.
Install the scale on a solid floor where there is no draft, vibration, strong magetic
fields or direct sunlight.
Consider section "4. Caution".
Step 7 Display normal weighing for at least 30 minutes to warm up the scale.