HVG-WP Series
Page 51
16.2.
Stream Mode
The scale outputs the current weighing data at the time when the disply is refreshed.
The scale does not output data while in the setting mode.
Averaging of function table
Refresh rate
A
f13 0
Fast response
Sensitive value
Approximately 7 times/s while the display is unstable,
Approximately 4 times/s when the display is stable
A
f13 1
c
A
f13 2
Slow response
Stable value
Approximately 4 times/s
A
16.2.1.
Preparation and Performance (Examples)
Step 1 Turn off the display.
Press and hold the
ZERO
key and press the
ON/OFF
key. The function table is displayed.
Step 2 Press the
ENTER
key to display
f4
(“Baud rate”).
Select a parameter of “Baud rate” with the
∧
keys.
Step 3 Press the
ENTER
key to display
f5
(“output
mode").
Select “Stream mode (
f5 0
)” with the
∧
keys.
Step 4 Press the
ENTER
key to display
f15
(“Format").
Select “Format 1 (
f15 0
)” with the
∧
keys
Step 5 Press the
ENTER
key to store the new parameters.
The scale returns the normal weighing and outputs
the data coutinuouslly.