Page 26
HV/HW-C/CP Series
Tare Function to Display Net Value
The
tare function
is used to cancel the mass value of a tare weight and to display the
net
value
, when a container (tare weight) to hold the article to be weighed is placed on the
weighing pan.
Caution
The weighing range is reduced according to the value of the tare weight.
The current net value is reset to zero value when the ZERO key is pressed or the display is
turned off.
For the HV
-
C and HV
-
CP series, the storable preset tare value must be within the minimum
weighing range.
7.2.1. Inputting Tare Value by Weighing
How to weigh a tare weight (container) and display the net value.
Step 1 Place the container on the weighing pan.
Wait for the stability mark to be displayed.
Press the TARE key.
The display becomes zero and the net mark is displayed.
Step 2 Place an article to be weighed in the container.
Wait for the stability mark to be displayed and read the net display.
Step 3 Remove all of the articles and the container from the weighing pan.
7.2.2. Digital Input of Tare Value (Preset Tare)
How to input value of a tare weight using numerical keys and to display the net value.
Step 1 While pressing and holding the SET key, press the TARE key.
The scale displays a blank display or the stored tare value. A blank display means that the
tare value is zero (reset value). The
PT mark blinks.
Step 2 Input the preset tare value using the numerical keys.
Step 3 Press the ENTER key to store the new preset tare value.
The scale displays the net value as preset tare value subtracted from gross value.
Step 4 Place an article to be weighed in the container.
Wait for the stability mark to be displayed and read the net display.
Step 5 Remove all of articles and the container from the weighing pan.