Page 44
HV/HW-C/CP Series
Calibration test report
The calibration test mode is used to compare a calibration weight
with the calibration test data weighed by the scale.
Note
This test does not perform calibration.
Step 1 In the weighing mode, press and hold the CAL switch
until
appears, and then release the switch.
The calibration test mode is not available when the
function setting
,
or
is selected.
Step 2 Press the ENTER key to display
.
Step 3 If necessary, change the value of the calibration weight
as described in "
14.2.4. Span Calibration
" .
Step 4 With nothing on the weighing pan, press the ENTER key.
The zero point is measured and the measured value with the
unit "kg" is displayed for a few seconds. Then, the value of
the calibration weight is displayed.
Step 5 Place a weight of the same value as displayed on the
weighing pan and press the ENTER key to measure
it. The measured value with the unit "kg" is displayed for
a few seconds.
Step 6
appears.
Step 7
appears and the calibration test report is output.
Step 8
appears again. Remove the weight.
Press the ON/OFF key to turn the display off or
press the CAL switch.
Press
for
5
Press
seconds
or
more
or
Release
key
Change mass value if necessary
Weighing
unit
Weighing
unit
Output
result