Page 70
HV/HW-C/CP Series
To output the number of accumulations.
Command
N C
R
L
F
Reply
N , + 0 0 0 0 0 1 4 0 C
R
L
F
To clear accumulation.
Accumulated data and number of accumulations are set to zero.
Command
C A C
R
L
F
Reply
C A C
R
L
F
To set the ID number.
The ID number is numerical value and alphabet characters of 6 digits and does not contain a
decimal point.
Template
I D
:
[ parameter ]
Command
I D
:
A B C 1 2 3 C
R
L
F
Reply
I D
:
A B C 1 2 3 C
R
L
F
To request the ID number.
The ID number is output.
Command
? I D C
R
L
F
Reply
I D
:
A B C 1 2 3 C
R
L
F
To output the preset tare value.
The preset tare value is output.
Command
? P T C
R
L
F
Reply
P T , + 0 0 0 0 , 1 2 0 k g C
R
L
F
To output H3 value.
Five-level
comparator
mode.............Not used
Three-level comparator mode..........Not used
Seven-level comparator mode .........Threshold value of rank 5 is output.
Command
? H 3 C
R
L
F
Reply
H 3 , + 0 0 0 5 0 0 C
R
L
F
To output H2 value.
Five-level
comparator
mode.............HIHI limit value is output.
Three-level comparator mode..........HI limit value is output.
Seven-level comparator mode .........Threshold value of rank 4 is output.
Command
? H 2 C
R
L
F
Reply
H 2 , + 0 0 0 4 0 0 C
R
L
F