HV-G Instruction Manual Page 42
14.2.3 Calibration of the Zero Point
Step 8 Press and hold the
CAL
switch
to enter the calibration mode
after displaying normal
weighing for 30 minutes.
Then the
Cal 0
is displayed.
Step 9 With nothing on the pan, press
the
ENTER
switch while the sta-
ble mark is displayed. The scale
stores the current condition as
the zero point. .
Step 10
The scale displays
5pn1
for
several seconds. If finishing
the calibration mode at this
stage (Zero calibration only),
proceed to step 14.
14.2.4 Span Calibration
Step 11 Set the value of the calibration
mass using the following switches.
(This initial value is according to
each product.)
^
switch
Selecting the value of
a digit.
<
switch
Selecting a digit.
Step 12 Place the mass’ selected at
step 11’ onto the pan and press
the
ENTER
switch while the
stable mark is displayed. The
scale then calculates the span
and stores it.
Step 13 The scale displays
end
at the
finish.
Remove the mass from the pan.
Step 14
Press the
CAL
switch to return
to the normal weighing mode.