HV-G Instruction Manual Page 51
16.2 Stream Mode
The scale outputs the current display data at the same time as refreshing the display.
The scale does not output data while in the setting mode.
Averaging function table
Refresh rate
f13 0
Normal weighing
Approximately 7 times/s while the display is
unstable.
Approximately 4 times/s when the display is stable
f13 1
A person’s weight
Approximately 4 times/s
f13 1
Animal weighing
Approximately 4 times/s
Step 1
Turn off the display.
Press the
ON/OFF
switch while the
ZERO
switch is
pressed and held. The function table is displayed.
Step 2
Select a parameter for the “baud rate (
f4
)”
using the
ENTER
and
^
switches.
Step 3
Select “stream mode (
f5 0
)” of “output” using
the
ENTER
and
^
switches.
Step 4
Select “format 1 (
f15 0
)” of “format” using the
ENTER
and
^
switches.
Step 5
Press the
ENTER
switch to store the new
parameters. The scale returns to normal weighing
and outputs the data countinuously.
16.2.1 Preparation and Performance (Examples)
Summary of Contents for HV-G Series
Page 1: ...H HV V G G S Se er ri ie es s P Pl la at tf fo or rm m S Sc ca al le es s Instruction Manual ...
Page 6: ...HV G Instruction Manual Page 5 3 Unpacking ...
Page 12: ...HV G Instruction Manual Page 11 6 Names ...
Page 67: ...HV G Instruction Manual Page 66 Dimensions ...