Page 18
HV-WP/HW-WP Series
7.2. Tare (And Net Display)
"Tare" is used to cancel the mass of a container, receptacle, case, bag, etc. which is
put on the weighing pan to contain the item to be weighed.
Caution
Using a tare value reduces the weighing range.
The current tare value is reset by pressing the
ZERO
switch or turning the
scale off. (Reset value is zero.)
The preset tare value must be within the minimum weighing range for the HV-WP series.
7.2.1. The Way of Tare Input by Weighing
Step 1 Put the container item on the weighing pan.
Step 2 Wait for the stability mark to be displayed. Press the
TARE
switch. The display
becomes zero and the net mark is displayed.
Step 3 It is now possible to put something into the container. Wait for the stability mark to
be displayed and to read its net display.
Step 4 Remove all things on the weighing pan.
7.2.2. The Way of Digital Input (Preset Tare)
Step 1 Press and hold the
SET
switch and press the
TARE
switch.
Then the blank or stored tare value is displayed. A blank display means that the
tare value is zero (reset value), and blinks.
Step 2 Set the preset tare value by using the following switches.
∧
switch
Selecting the number of the figure.
<
switch
Selecting a figure.
Step 3 Press the
ENTER
switch to store the new preset tare value.
Then the scale displays a net value with the tare value subtracted from the gross
weighing value.
Step 4 It is then possible to put something into the container. Wait for the stability mark to
be displayed and to read its net.
Step 5 Remove all things from the weighing pan.
7. Basic Operation
Summary of Contents for HW-200KV-WP
Page 72: ...Page 70 HV WP HW WP Series MEMO...