Page 26
HV-WP/HW-WP Series
10. Accumulation Function
10. Accumulation Function
This function counts the number of weighed items, calculates the total mass value
and can display the number and accumulated mass value.
The accumulation function is displayed with up to 6 digits. The balance can not display 7 or
more digits, therefore the leading digits are not displayed.
Example: With 60K type, when importing the data of 17 accumulations of up to 60kg (60.000
X 17=1020.000), the balance displays this as “
020.000
”.
Set the parameters of the "accumulation function (
f8
)" in the function table, in
advance, to use this function.
The accumulation count and accumulated mass value are stored in the scale even if
the power is removed.
Operation and Switches
The display of the accumulation count has a blinking without a unit.
The display of the accumulated mass value has a unit and a blinking .
Pressing the
MODE
switch, the accumulation count and accumulated mass value are
displayed.
Pressing
the
ZERO
switch in the accumulation function ( is blinking), the current
function resets. (The count and accumulated mass value become zero.)
Caution
The accumulation function can be used with the first weighing unit accumulated.
Parameter List and Word Definition
The "nearly-zero" is within
±
5 digits from the zero point in the unit of kg.
The "digit", a unit of display, is equivalent to the minimum measurable mass.
The "zero point" is the fundamental starting point to weigh anything.
Function table
Meaning and purpose
A
f8 0
Accumulation function not used.
A
f8 1
The scale accumulates the data, if the
F
switch is pressed, when
the display is a positive stable value without nearly-zero. The next
accumulation can be performed after the display is nearly-zero or a
negative value.
A
A
A
A
f8 2
The scale accumulates the data, if the
F
switch is pressed, when
the display is a stable value and without nearly-zero. The next
accumulation can be performed after the display is nearly-zero.
A
A
A
f8 3
When the display is a positive stable value and without nearly-zero,
the scale accumulates the data automatically. The next
accumulation can be performed after the display is nearly-zero or a
negative value.
A
A
A
Summary of Contents for HW-200KV-WP