20
MS-70 / MX-50 / MF-50 / ML-50
1. Turn on the analyzer.
The gram unit (of weighing mode) is displayed.
Enter the Standard Mode
2. Press
the
SELECT
key to display an analysis
mode and press the
↓
or
↑
key to select
5td
.
Select ACCURACY
3. Press
the
SELECT
key to select
ACCURACY.
When
ACCURACY
is selected, HI, MID. or LO blinks.
4. Press
the
↓
or
↑
key to select LO of ACCURACY.
Set Drying Temperature at the Sample Pan
5. Press
the
SELECT
key to select the drying
temperature.
6. Press
the
↓
or
↑
key to set 130 °C.
Store the Parameters and Finish the Operation
7. Press
the
ENTER
key to store the parameters.
The weighing mode is automatically displayed.
Put a Sample on the Pan
8. Put the breeze break ring, pan support, pan handle
and sample pan in order. (With no sample.)
9. Close the heater cover.
10. When displaying a stable value, Press the
RESET
key. Avoid mechanical vibration, breeze and
environmental noise during measurement. If it
deviates from zero display, press the
RESET
key.
Close
Preparation
Standard mode
ACCURACY
Temperature
Standard drying
ML-50 does not display