54
MS-70 / MX-50 / MF-50 / ML-50
11. Data Memory Function
The data memory function automatically stores each result when finishing a measurement.
MS-70
/
MX-50
MF-50
ML-50
Maximum number
100 data
50 data
30 data
The stored data can be output to a printer at one time and can be output to a computer
using RsCom and Rskey that are Windows applications stored in the CD-ROM of
WinCT-Moisture or WinCT at one time.
The stored data can be deleted at one time.
The function can select either storing each result or not at
data
of the function table.
Data is stored at each measurement..........
data 1
Data is not stored .......................................
data 0
When using the data memory function,
MEM
is displayed.
When displaying
fUll
, the function can not store the next data. The function can
store new data after deleting the stored data.
Caution
When pressing the
STOP
key during a measurement except manual mode,
the result is not stored.
Set
data 1
before measurement, if it is necessary to store each result with data
memory function.
11.1.1. Preparation
This example selects "store result" at
data
of the function table.
1. Display the gram unit (of the weighing mode).
2. Press and hold the
SELECT
key to enter the
function table.
3. Press
the
SELECT
key to display
data
.
3. Press
the
SELECT
key several times and press
the
↓
or
↑
key to display
data "1
.
5. Press
the
ENTER
key to store the new settings.
Press
the
RESET
key to return to the weighing mode.
MEM is displayed when the memory function is effective.
Memory function sign
Number of data
Full memory
0.000
g
Cl
adj
data "1
Press and hold
data
Store result
5-d
MEM
0
0.000
g