AND MX-50/MF-50 Instruction Manual
GB 36
S p e c i f i c a t i o n s
14
MX-50
MF-50
Measurement method
400 W Halogen lamp
Drying temperature range at sample pan
50 °C to 200 °C (1 °C increments)
Temperature calibration
By Accessory AX-MX-43
---
Sample weight range
0.1 g to 51 g
Accuracy: Repeatability of measurement, (Standard deviation)
Moisture
over 5 g sample
0.02%
0.05%
content *
1
over 1 g sample
0.1%
0.2%
Weighing mode
0.001g
0.002g
Minimum reading
Moisture content
0.01%, 0.1%
0.05%, 0.1%, 1%
Weighing mode
0.001g
0.002g
Measurement programs
Analysing
Standard mode
Set accuracy from HI, MID or LO and select drying temperature.
mode
Sample weight and termination value is set automatically.
When the drying rate reaches fixed termination value,
measurement is automatically completed. (*
2
)
Automatic
When drying rate is less than preset termination value,
complete mode measurement is automatically completed. (*
2
)
Timer mode
After heating at the preset drying time, measurement is
automatically stopped. (1min. to 480 min.)
Manual mode When pressing the key at any time, measurement is stopped
and the result is decided.
Measurement unit
Moisture content (Wet-base)
Moisture content (Dry-base, Atro)
Solid content
Ratio (%)
Number of memory
20 set
10 set
Data memory function
Storage of 100 results
Storage of 50 results
Communication function
RS-232C serial interface
Application software for Windows (CD-ROM) WinCT-Moisture Analysing utility
WinCT Communication software
Operation environment
5 °C to 40 °C (41°F to 104°F), 85%RH or less (no condensation)
Sample pan
85 mm diameter
Power source, Maximum current (rms), AC 200 V to 240 V, 1.5A
Maximum consumption
50Hz or 60Hz, Approximately 400W
External dimensions
215(W) x 320(D) x 173(H)mm,
Mass (Net weight)
Approximately 6kg (without accessories)
Housing material
Flame-retardant (UL94V0)
*
1
: After preheating the analyser, the data can be obtained with approximately 5 g calibration sample
(Sodium tartrate dihydrate) in standard mode at 160 °C
*
2
: When the drying rate reaches to the preset termination value, the measurement is completed.
User mode