3. Setting Command – s**
Communication for RA2000 series (1WMPD4003507)
3-5
SML (Set Memory Length) Setting memory block size
SMB (Set Memory Block) Setting block No.
STD (Set Trigger Delay) Setting pre-trigger
STE (Set Trigger Execution) Setting trigger execution
SMC(Set Memory Copy) Sets the readout amount
For the other settings, see commands in the following table.
Function
Sets the memory block size using arbitrary values.
SML P1 (Delimiter)
Input Format
P1: Data count stored in memory (1000 to 61865984)
* The effective size (upper limit value) is a value determined by dividing
memory capacity by channel count.
* The upper limit value of P1 above is a value when the number of data
recording channel is one. In actual cases, the upper limit value will be the
value divided by the number of recording channel (set by SRC Recording
Channel).
Output Format None
Description
While the RA2300MK
Ⅱ
is operating, an execution error occurs.
When the memory block size is changed, the memory data is deleted.
Memory segmentation up to 128 is made, which is larger segmentation than
specified size.
Function
Setting block No..
SMB P1 (Delimiter)
Input Format
P1: Block No. ([1 - 128])
Output Format None
Description
While the RA2000 is operating, an execution error occurs.
The range varies depending on the segmentation number.
(Example: 8 segmentation, [1-8])
Function
Sets pre-trigger.
STD P1 (Delimiter)
Input Format
P1: Pre-trigger ([0-100]%)
Output Format None
Description
While the RA2000 is operating, an execution error occurs.
Becomes valid when recording in a memory block.
Function
Sets trigger execution.
STE P1 (Delimiter)
Input Format
P1: Trigger execution (1=Once, 2=Repeat, 3=Endless)
Output Format None
Description
While the RA2000 is operating, an execution error occurs.
Be reflected only in memory recording.
Function
Sets the readout amount of the internal memory when copying
SMC P1(Delimiter)
Input Format
P1: readout amount(1-100 %)
Output Format None
Description
While the RA2000 is operating, an execution error occurs.
Setting contents
Command to see
Path to save files
SSS (Set filing Save Setting) Setting place where to save files
Setting for CSV savings
SMF (Set Memory Filing) Setting Filing
Summary of Contents for Omniace RA2000 series
Page 1: ...Communication Command RA2000 series INSTRUCTION MANUAL 1WMPD4003507 ...
Page 2: ......
Page 3: ...Communication Command RA2000 series INSTRUCTION MANUAL ...
Page 4: ......
Page 14: ......
Page 15: ...1 Selection of Communication Interface ...
Page 27: ...2 Overview of Communication Control ...
Page 34: ......
Page 35: ...3 Setting Command s ...
Page 63: ...4 Information Readout Command I ...
Page 90: ......
Page 91: ...5 Execution Command E ...
Page 99: ...6 File Data Operation Command F ...
Page 102: ......
Page 103: ...7 Text Operation Command T ...
Page 109: ...8 Reference ...
Page 113: ......