3. Setting Command – s**
Communication for RA2000 series (1WMPD4003507)
3-6
3.5. HD Recording
SRF (Set Real-time Filing) Setting HD recorder basics
SFT (Set Filing Time) Setting recording time
SRT (Set Real-time Trigger) Setting real-time recording operation
For the other settings, see commands in the following table.
Function
Sets recording speed, recording length, and recording method.
SRF P1, P2, P3, P4, P5 (Delimiter)
Input Format
P1: Recording speed numeric value
P2: Recording speed Unit
P3: Data format
P4: Recording method
P5: Recording data number
([1-1000, E] E=external synchronization)
(1=[
μ
s], 2=[ms], 3=[s]) Invalid when P1=E.
(1=Peak, 2=Sampling)
(1=Normal, 2=Ringing)
(Selecting 0 enables the whole “Free Disk
Space”)
Output Format None
Description
While the RA2300MK
Ⅱ
is operating, an execution error occurs.
The recording speed settings with P2 are: 1
s – 100s for RA2300MK
Ⅱ
, and
1
s – 100s for RA2800A.
The recording speed settings with P1 and P2 are limited from 1us to 10s.
Function
Sets recording time.
SFT P1, P2, P3, P4 (Delimiter)
Input Format
P1: Day number (0 or higher numeric value) (To be omitted, select 0)
P2: Hour number (0 or higher numeric value) (To be omitted, select 0)
P3: Minute number (0 or higher numeric value) (To be omitted, select 0)
P4: Second number (0 or higher numeric value) (To be omitted, select 0)
Output Format None
Description
While the recorder is operating, an execution error occurs.
P1 through P4 can be omitted. When the value is omitted, 0 is set. However, if
all the values are omitted, a parameter error occurs.
Function
Sets real-time recording operation.
SRT P1, P2 (Delimiter)
Input Format
P1: Starting execution of recording with detecting trigger.
P1
Starting execution of recording with detecting trigger
0
Pressing “START” key initiates recording soon.
1
Detecting trigger initiates recording.
2
Detecting trigger initiates and repeats recording.
P2: Mark printing with trigger (0=OFF, 1=ON)
Output Format None
Description
While the RA2300MK
Ⅱ
is operating, an execution error occurs.
The repeat execution with P1=2 is valid only when a recording length is
limited.
Setting contents
Command to see
Paper feeding speed of a
wavelength chart recording
SCS (Set Chart Speed) Setting paper feed speed of waveform chart
Path to save files
SSS (Set filing Save Setting) Setting place where to save files
Summary of Contents for Omniace RA2000 series
Page 1: ...Communication Command RA2000 series INSTRUCTION MANUAL 1WMPD4003507 ...
Page 2: ......
Page 3: ...Communication Command RA2000 series INSTRUCTION MANUAL ...
Page 4: ......
Page 14: ......
Page 15: ...1 Selection of Communication Interface ...
Page 27: ...2 Overview of Communication Control ...
Page 34: ......
Page 35: ...3 Setting Command s ...
Page 63: ...4 Information Readout Command I ...
Page 90: ......
Page 91: ...5 Execution Command E ...
Page 99: ...6 File Data Operation Command F ...
Page 102: ......
Page 103: ...7 Text Operation Command T ...
Page 109: ...8 Reference ...
Page 113: ......