3. Setting Command – s**
Communication for RA2000 series (1WMPD4003507)
3-8
3.7. Trigger
STM (Set Trigger Mode) Setting trigger mode
STC (Set Trigger mode OR, AND Channel) Setting OR, AND trigger
condition
STW (Set Trigger Window) Setting WINDOW trigger condition
Function
Sets trigger mode.
STM P1, P2 (Delimiter)
Input Format
P1: Trigger mode 0=OFF, 1=OR, 2=AND, 4=WINDOW
P1
Trigger mode
0
OFF (memory block=1 block)
1
OR
2
AND
3
<Reserved> A parameter error occurs.
4
WINDOW
P2: <Reserved>
Output Format None
Description
While the RA2300MK
Ⅱ
is operating, an execution error occurs.
The recorder does not support P1=3(A*B); therefore, a parameter error
occurs when it is selected.
Function
Sets OR, AND trigger condition.
STC P1, P2, P3, P4 (Delimiter)
P1: Channel number [1-16] for RA2300MK
Ⅱ
and [1-32] for RA2800
P2: Detecting ON/OFF 0=OFF, 1=ON
P3: Varies depending on amp type (see below). (Can be omitted.)
P4: Varies depending on amp type (see below). (Can be omitted.)
----- For analog type of amp -----
P3: Trigger level
Selecting with measured value
(within the dynamic range).
P4: Slope 1=Rising edge, 2=Falling edge)
Input Format
----- For event amp -----
P3: Detecting logic 1=AND, 2=OR
P4: Detecting pattern 0=X, 1=H, 2=L
Example: For HHLL XXHL, “11220012”.
Output Format None
Description
P1=E1 is a setting for an extra event or the recorder event (E1).
While the RA2300MK
Ⅱ
is operating, an execution error occurs.
When the selected channel is an invalid amp, a parameter error occurs.
Function
Sets WINDOW trigger condition.
STW P1, P2, P3, P4, P5, P6 (Delimiter)
Input Format
P1: Channel number [1-16] for RA2300MK
Ⅱ
and [1-32] for RA2800A
P2: Detecting ON/OFF
0=OFF, 1=ON
P3: <Reserved>
P4: Maximum trigger level
Selecting with measured value (within the dynamic range).
P5: Minimum trigger level
Selecting with measured value (within the dynamic range).
P6: Trigger occurrence direction 1=IN, 2=OUT
Output Format
None
Description
While the RA2300MK
Ⅱ
is operating, an execution error occurs.
When the selected channel is the amp other than an analog type of amp, a
parameter error occurs.
Summary of Contents for Omniace RA2000 series
Page 1: ...Communication Command RA2000 series INSTRUCTION MANUAL 1WMPD4003507 ...
Page 2: ......
Page 3: ...Communication Command RA2000 series INSTRUCTION MANUAL ...
Page 4: ......
Page 14: ......
Page 15: ...1 Selection of Communication Interface ...
Page 27: ...2 Overview of Communication Control ...