3. Setting Command – s**
Communication for RA2000 series (1WMPD4003507)
3-11
SCH (Set CHannel) Setting FFT amp
SCH (Set CHannel) Setting HSDC amp
Function
Setting FFT amp.
SCH P1, P2, P3, P4, P5, P6, P7, to P13 (Delimiter)
P1: Specifying channel
P2: amp type
P3: Input
P4: Setting range
P5: Filter
P6: Position
P7: Input combination
P8: Measurement mode
P9: Setting sensor
P10: Vibration unit
P11: Hybrid-type sensor
sensitivity
P12: Charge converter
sensitivity
P13: Acceleration
Sensor sensitivity
[1-16, A] for RA2300MK
Ⅱ
and [1-32, A] for
RA2800A A means a batch setting.
2 fixed
0=OFF, 1=ON, 2=GND
1=500V, 2=200V, 3=100V, 4=50V, 5=20V,
6=10V, 7=5V, 8=2V, 9=1V, 10=500mV,
11=200mV, 12=100mV
0=OFF, 1=30Hz, 2=300Hz, 3=3kHz,
4=Anti-aliasing
[-100.00 to 200.00] Step 0.05
1=AC, 2=DC
0=Voltage, 1=Vibration
1=Hybrid type, 2=Standalone type
1=[m/s^2], 2=[G]
[0.001 to 120.000]mV/m/s^2 or
[0.010 to 1200.00]mV/G
[0.01 to 10.0]mV/pC
[0.001 to 120.000]pC/m/s^2 or
[0.010 to 1200.00]pC/G
Input Format
The sensitivity ranges of P11 and P13 vary depending on a vibration unit.
Output Format
None
Description
When P1 = A, the other selections apply to all the channels corresponding
to the selected type with P2.
When the amp type of selected channel does not correspond to P2, a
parameter error occurs.
While any action other than a chart recording is executing, an execution
error occurs.
When a measurement mode is vibration (P8=1), the setting range of P4 is
5V-100mV (7-12).
Function
Sets HSDC amp.
SCH P1, P2, P3, P4, P5, P6, P7 (Delimiter)
Input Format
P1: Selecting channel
P2: Amp type
P3: Input
P4: Setting range
P5: Filter
P6: Position
P7: Input combination
[1-16, A] for RA2300MK
Ⅱ
and [1-32, A] for
RA2800A A means a batch setting.
3 fixed
0=OFF, 1=ON, 2=GND
1=500V, 2=200V, 3=100V, 4=50V, 5=20V,
6=10V, 7=5V, 8=2V, 9=1V, 10=500mV,
11=200mV, 12=100mV
0=OFF, 1=5Hz, 2=50Hz, 3=500kHz, 4=5kHz,
5=50kHz
[-100.00 to 200.00] Step 0.05
1=AC, 2=DC
Output Format
None
Description
When P1 = A, the other selections apply to all the channels corresponding
to the selected type with P2.
When the amp type of selected channel does not correspond to P2, a
parameter error occurs.
While any action other than a chart recording is executing, an execution
error occurs.
Summary of Contents for Omniace RA2000 series
Page 1: ...Communication Command RA2000 series INSTRUCTION MANUAL 1WMPD4003507 ...
Page 2: ......
Page 3: ...Communication Command RA2000 series INSTRUCTION MANUAL ...
Page 4: ......
Page 14: ......
Page 15: ...1 Selection of Communication Interface ...
Page 27: ...2 Overview of Communication Control ...
Page 34: ......
Page 35: ...3 Setting Command s ...
Page 63: ...4 Information Readout Command I ...
Page 90: ......
Page 91: ...5 Execution Command E ...
Page 99: ...6 File Data Operation Command F ...
Page 102: ......
Page 103: ...7 Text Operation Command T ...
Page 109: ...8 Reference ...
Page 113: ......