3. Setting Command – s**
Communication for RA2000 series (1WMPD4003507)
3-16
SCH (Set CHannel) Setting DCST amp
SCH (Set CHannel) Setting HRZS amp
Function
Sets DCST amp.
SCH P1, P2, P3, P4, P5, P6, P7, P8 (Delimiter)
P1: Selecting channel
P2: Amp type
P3: Input
P4: Setting range
[1-16, A] for RA2300MK
Ⅱ
and [1-32, A] for
RA2800A A means a batch setting.
10 fixed
0=OFF, 1=ON, 2=GND
The content varies depending on the P8 Input
mode.
P8=1 ST BV=2V
1=50k
με
, 2=20k
με
, 3=10k
με
, 4=5k
με
, 5=2k
με
P8=2 ST BV=5V
1=20k
με
, 2= 8k
με
, 3= 4k
με
, 4=2k
με
, 5=800
με
P8=3 DC
1=50mV, 2=20mV, 3=10mV, 4=5mV, 5=2mV
Input Format
P5: Filter
P6: Position
P7: Gage rate
P8: Input mode and BV
0=OFF, 1=10Hz, 2=30Hz, 3=300Hz, 4=1kHz
[-100.00 to 200.00] Step 0.05
[1.50 to 2.50] Step 0.01 Select at 2.00 for out
of range
1=ST(BV=2V), 2=ST(BV=5V), 3=DC
Output Format
None
Description
When P1 = A, the other selections apply to all the channels corresponding
to the selected type with P2.
When the amp type of selected channel does not correspond to P2, a
parameter error occurs.
While any action other than a chart recording is executing, an execution
error occurs.
“BV” means a bridge voltage.
Function
Sets HRZS amp.
SCH P1, P2, P3, P4, P5, P6, P7, P8, P9 (Delimiter)
P1: Selecting channel
P2: Amp type
P3: Input
P4: Setting range
P5: Filter
P6: Position
P7: Input combination
P8: ZSV ON/OFF
P9: ZSV level
[1-16, A] for RA2300MK
Ⅱ
and [1-32, A] for
RA2800A A means a batch setting.
11 fixed
0=OFF, 1=ON, 2=GND
1=500V, 2=200V, 3=100V, 4=50V, 5=20V, 6=10V,
7=5V, 8=2V, 9=1V, 10=500mV, 11=200mV,
12=100mV
0=OFF, 1=30Hz, 2=300Hz, 3=3kHz
[-100.00 to 200.00] Step 0.05
1=AC, 2=DC
0=OFF, 1=ON
The range varies depending on a P4 range
setting.
500V-5V:[-130.000 to 130.000]V
2V-100mV:[-13.0000 to 13.0000]V
Input Format
ZSV means a zero suppression voltage.
Output Format
None
Description
When P1 = A, the other selections apply to all the channels corresponding
to the selected type with P2.
When the amp type of selected channel does not correspond to P2, a
parameter error occurs.
While any action other than a chart recording is executing, an execution
error occurs.
Summary of Contents for Omniace RA2000 series
Page 1: ...Communication Command RA2000 series INSTRUCTION MANUAL 1WMPD4003507 ...
Page 2: ......
Page 3: ...Communication Command RA2000 series INSTRUCTION MANUAL ...
Page 4: ......
Page 14: ......
Page 15: ...1 Selection of Communication Interface ...
Page 27: ...2 Overview of Communication Control ...
Page 34: ......
Page 35: ...3 Setting Command s ...
Page 63: ...4 Information Readout Command I ...
Page 90: ......
Page 91: ...5 Execution Command E ...
Page 99: ...6 File Data Operation Command F ...
Page 102: ......
Page 103: ...7 Text Operation Command T ...
Page 109: ...8 Reference ...
Page 113: ......