4. Information Readout Command - I**
Communication for RA2000 series (1WMPD4003507)
4-6
IMS (Inquire Memory Status) Read-out of memory status
Function
Outputs the status of memory.
Depending on the parameter, output format will vary.
The information of a memory block, which was set by a monitor or a SMB
command at present, will output.
Input Format
IM S P1 (Delimiter)
P1: Point out an information category.
(Omit able P1 is as same as response 0)
P1 Information category
0
Output existence of data
※
1
1
Output of the Start, Trigger, and End time
※
2
2
Output existences of all data among 128 blocks
※
1
3
Output of Data existence, start, trigger, end time
※
2
4
Output of Trigger, end address
※
2
5
Output of maximum block No in which all the data are effective.
※
1
P1 =0: A1 (Delimiter) (Same at the time of P1 omission):
0= no data, 1= data exists
In case of P1 =1: A1, A2, A3 (Delimiter)
A1 = Starting time of recording, A2 = Trigger detection time, A3 = Ending
time of recording
Fromat of time indication: YY/MM/DD HH:MM:SS
No trigger case: A2="**/**/** **:**:**"
No data case A1=A2=A3="**/**/** **:**:**"
In case of P1 =2: A1, A2, A3, ......, A128 (Delimiter)
An Block
Data
0
Exist None
1
Exist Exist
*
None None
Example: If the case of four even splits, only data exist in block 1
A1="1",A2=A3=A4="0", A5 ~A128="*"
In case of P1 =3: A1, A2, A3, A4 (Delimiter)
A1 is the same as A1 in case of (P1 =0)
A2, A3, A4 are the same as A1, A2, A3 in case of (P1 =1)
In case of P1 =4: A1, A2 (Delimiter)
A1 outputs a trigger address(numerical value). If there is no detection, A1
outputs "*".
A2 outputs a last address(numerical value). If there is no data, A1 outputs "*".
Output Format
In case of P1 =5: A1(Delimiter)
Data output the greatest number of an effective block.
If there is no data which is effective for a block, it outputs "*".
Description
The reference of memory block out of case of P1-2 or 5 will be present active
block. (Point the setting of “SMB block No”)
If the case P1 is not correct, it becomes a parameter error. This case of
answer output is “?”.
Summary of Contents for Omniace RA2000 series
Page 1: ...Communication Command RA2000 series INSTRUCTION MANUAL 1WMPD4003507 ...
Page 2: ......
Page 3: ...Communication Command RA2000 series INSTRUCTION MANUAL ...
Page 4: ......
Page 14: ......
Page 15: ...1 Selection of Communication Interface ...
Page 27: ...2 Overview of Communication Control ...
Page 34: ......
Page 35: ...3 Setting Command s ...
Page 63: ...4 Information Readout Command I ...
Page 90: ......
Page 91: ...5 Execution Command E ...
Page 99: ...6 File Data Operation Command F ...
Page 102: ......
Page 103: ...7 Text Operation Command T ...
Page 109: ...8 Reference ...
Page 113: ......