4. Information Readout Command - I**
Communication for RA2000 series (1WMPD4003507)
4-9
4.7. Trigger
ITM (Inquire Trigger Mode) Reading trigger mode
ITC (Inquire Trigger mode OR,AND Channel) Reading OR, AND trigger
condition
ITW (Inquire Trigger Window) Reading WINDOW trigger condition
ITF (Inquire Trigger Filter) Reading trigger filter
Function
Outputs trigger mode setting.
Input Format ITM (Delimiter)
A1 (Delimiter)
Output Format
A1: Trigger Mode 0=OFF, 1=OR, 2=AND, 4=WINDOW
Description
The recorder does not support A1=3(A*B).
Function
Outputs the setting of OR, AND trigger condition.
ITC P1 (Delimiter)
Input Format
P1: Channel number (For RA2300MK
Ⅱ
, 1 through 16, for RA2800A, 1
through 32)
A1, A2, A3 (Delimiter)
A1: Detecting ON/OFF 0=OFF, 1=ON
A2: Varies depending on amp type (see below).
A3: Varies depending on amp type (see below).
----- For analog type of amp -----
A2: Trigger level Represents with the measurement value.
A3: Slope 1=Rising edge, 2=Falling edge
Output Format
----- For event amp -----
A2: Detecting logic 1=AND, 2=OR
A3: Detecting pattern 0=X, 1=H, 2=L
Outputs Sig1, Sig2, to Sig8 in the order from left.
Example: For HHLL XXHL, “11220012”.
Description
When the selected channel is an invalid amp, a parameter error occurs.
When an error occurs, "?,?,?" is returned.
Function
Outputs the setting of WINDOW trigger condition.
ITW P1 (Delimiter)
Input Format
P1: Channel number
(For RA2300MK
Ⅱ
, 1 through 16 and E1, for RA2800A, 1 through 32, and
E1)
A1, A2, A3, A4, A5 (Delimiter)
Output Format
A1: Detecting ON/OFF
A2: <Reserved>
A3: Maximum trigger level
A3: Minimum trigger level
A5: Trigger occurrence direction
0=OFF, 1=ON
Represents with the measurement value.
Represents with the measurement value.
1=IN, 2=OUT
Description
When the selected channel is an invalid amp, a parameter error occurs.
Function
Outputs trigger filter setting.
Input Format ITM (Delimiter)
A1 (Delimiter)
Output Format
A1: Trigger Filter [0-65534] 0=OFF
Description
Summary of Contents for Omniace RA2000 series
Page 1: ...Communication Command RA2000 series INSTRUCTION MANUAL 1WMPD4003507 ...
Page 2: ......
Page 3: ...Communication Command RA2000 series INSTRUCTION MANUAL ...
Page 4: ......
Page 14: ......
Page 15: ...1 Selection of Communication Interface ...
Page 27: ...2 Overview of Communication Control ...
Page 34: ......
Page 35: ...3 Setting Command s ...
Page 63: ...4 Information Readout Command I ...
Page 90: ......
Page 91: ...5 Execution Command E ...
Page 99: ...6 File Data Operation Command F ...
Page 102: ......
Page 103: ...7 Text Operation Command T ...
Page 109: ...8 Reference ...
Page 113: ......