4. Information Readout Command - I**
Communication for RA2000 series (1WMPD4003507)
4-10
4.8. Amp Unit
Names of input units are represented by the following symbols.
Name of Amp Unit
Symbol Name of Amp Unit
Symbol
2-CH high resolution DC amp unit
HRDC
TC/DC amp unit
TDC
2-CH FFT amp unit
FFT
F/V converter unit
FV
2-CH high speed DC amp unit
HSDC
2-CH oscillation
・
RMS amp unit
RMS
2-CH AC strain amp unit
ACST
2-CH DC strain amp unit
DCST
Event amp unit
EV
2-CH zero suppression amp unit
HRZS
2-CH TC/DC amp unit
TCDC
The decimal of the answer from the CH command is as follows.
Item
Decimal digit Item
Decimal digit
Position 2
Gauge
factor
2
Sensitivity of incorporated
sensor
3 Event
waveform
1
Charge converter sensitivity
2
ZS V (500 V-5 V)
3
Acceleration sensor sensitivity
3
ZS V (2 V-100m V)
4
ICH (Inquire CHannel) Reading HRDC amp Setting
Function
Outputs HRDC amp setting.
ICH P1 (Delimiter)
Input Format
P1: Selecting channel
(For RA2300MK
Ⅱ
, 1 through 16, for RA2800A, 1 through 32)
A1, A2, A3, A4, A5, A6 (Delimiter)
Output Format
A1: Amp type
A2: Input
A3: Setting range
A4: Filter
A5: Position
A6: Input combination
1 fixed
0=OFF, 1=ON, 2=GND
1=500V, 2=200V, 3=100V, 4=50V, 5=20V, 6=10V,
7=5V, 8=2V, 9=1V, 10=500mV, 11=200mV,
12=100mV
0=OFF, 1=30Hz, 2=300Hz, 3=3kHz
[-100.00 to 200.00] Step 0.05
1=AC, 2=DC
Description
Summary of Contents for Omniace RA2000 series
Page 1: ...Communication Command RA2000 series INSTRUCTION MANUAL 1WMPD4003507 ...
Page 2: ......
Page 3: ...Communication Command RA2000 series INSTRUCTION MANUAL ...
Page 4: ......
Page 14: ......
Page 15: ...1 Selection of Communication Interface ...
Page 27: ...2 Overview of Communication Control ...
Page 34: ......
Page 35: ...3 Setting Command s ...
Page 63: ...4 Information Readout Command I ...
Page 90: ......
Page 91: ...5 Execution Command E ...
Page 99: ...6 File Data Operation Command F ...
Page 102: ......
Page 103: ...7 Text Operation Command T ...
Page 109: ...8 Reference ...
Page 113: ......