4. Information Readout Command - I**
Communication for RA2000 series (1WMPD4003507)
4-11
ICH (Inquire CHannel) Reading FFT amp setting
ICH (Inquire CHannel) Reading HSDC amp setting
Function
Outputs FFT amp setting.
ICH P1 (Delimiter)
Input Format
P1: Selecting channel
(For RA2300MK
Ⅱ
, 1 through 16, for RA2800A, 1 through 32)
A1, A2, A3, A4, A5, A6 to A12 (Delimiter)
A1: Amp type
A2: Input
A3: Setting range
A4: Filter
A5: Position
A6: Input combination
A7: Measurement mode
A8: Setting sensor
A9: Vibration unit
A10: Hybrid-type sensor sensitivity
A11: Charge converter sensitivity
A12: Acceleration sensor sensitivity
2 fixed
0=OFF, 1=ON, 2=GND
1=500V, 2=200V, 3=100V, 4=50V,
5=20V, 6=10V7=5V, 8=2V,
9=1V, 10=500mV, 11=200mV,
12=100mV
0=OFF, 1=30Hz, 2=300Hz, 3=3kHz,
4=Anti-aliasing
[-100.00 to 200.00] Step 0.05
1=AC, 2=DC
0=Voltage, 1=Oscillation
1=Hybrid type, 2=Standalone type
1=[m/s^2],2=[G]
[0.001 to 120.000]mV/m/s^2 or [0.010
to 1200.00]mV/G
[0.01 to 10.0]mV/pC
[0.001 to 120.000]pC/m/s^2 or [0.010 to
1200.00]pC/G
Output Format
The sensitivity ranges of P11 and P13 vary depending on a vibration unit.
Description
Function
Outputs HSDC amp setting.
ICH P1 (Delimiter)
Input Format
P1: Selecting channel
(For RA2300MK
Ⅱ
, 1 through 16, for RA2800A, 1 through 32)
A1, A2, A3, A4, A5, A6 (Delimiter)
Output Format
A1: Amp type
A2: Input
A3: Setting range
A4: Filter
A5: Position
A6: Input combination
3 fixed
0=OFF, 1=ON, 2=GND
1=500V, 2=200V, 3=100V, 4=50V, 5=20V, 6=10V,
7=5V, 8=2V, 9=1V, 10=500mV, 11=200mV,
12=100mV
0=OFF, 1=5Hz, 2=50Hz, 3=500kHz, 4=5kHz,
5=50kHz
[-100.00 to 200.00] Step 0.05
1=AC, 2=DC
Description
Summary of Contents for Omniace RA2000 series
Page 1: ...Communication Command RA2000 series INSTRUCTION MANUAL 1WMPD4003507 ...
Page 2: ......
Page 3: ...Communication Command RA2000 series INSTRUCTION MANUAL ...
Page 4: ......
Page 14: ......
Page 15: ...1 Selection of Communication Interface ...
Page 27: ...2 Overview of Communication Control ...
Page 34: ......
Page 35: ...3 Setting Command s ...
Page 63: ...4 Information Readout Command I ...
Page 90: ......
Page 91: ...5 Execution Command E ...
Page 99: ...6 File Data Operation Command F ...
Page 102: ......
Page 103: ...7 Text Operation Command T ...
Page 109: ...8 Reference ...
Page 113: ......